NEOPUFF INFANT RESUSCITATOR
Report
- Report Number
- 9611451-2011-00265
- Event Type
- Malfunction
- Date Received
- April 26, 2011
- Date of Event
- March 25, 2011
- Report Date
- March 28, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTL
- PMA / PMN Number
- K892885
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). METHOD: THE COMPLAINT MANOMETER OF (B)(4) NEOPUFF INFANT RESUSCITATOR WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4). IT WAS VISUALLY INSPECTED FOR DAMAGE AND PERFORMANCE TESTED FOR FUNCTIONALITY. RESULTS: NO PHYSICAL DAMAGES WERE OBSERVED ON THE RETURNED MANOMETER. WHEN THE MANOMETER WAS ATTACHED TO A TEST VALVE ASSEMBLY, THE NEEDLE WAS STUCK AT 16 CMH2O. AN ATTEMPT WAS MADE TO ADJUST THE MANOMETER SETTING TO ZERO BUT IT COULD NOT BE RESET. CONCLUSION: THE FAULT OBSERVED WAS DUE TO THE MISALIGNMENT OF THE MANOMETER INTERNAL MECHANISM. THE NEOPUFF IS A PORTABLE, REUSABLE DEVICE USED TO ASSIST IN THE DELIVERY OF RESPIRATORY BREATHS TO INFANTS UNTIL ADEQUATE SPONTANEOUS BREATHING OCCURS. THE OPERATING MANUAL INSTRUCTS THE USER TO CARRY OUT A SET-UP PROCEDURE "PRIOR TO EVERY USE OF THE NEOPUFF TO ENSURE THAT THE DEVICE IS FUNCTIONING CORRECTLY". THE SET-UP PROCEDURE STATES THE USER MUST CHECK THE SETTINGS BY TESTING THE PRESSURE OUTPUT FROM THE NEOPUFF AT THE DESIRED FLOW RATE. THE NEOPUFF TECHNICAL MANUAL ALSO STATES THE FOLLOWING: "THE INTEGRITY OF THE SYSTEM AND MANOMETER SHOULD BE CHECKED PRIOR TO FIRST USE, ANNUAL AND AFTER SERVICING". "WARNING DROPPING OF THE F&P NEOPUFF INFANT RESUSCITATOR OR OTHER SIMILAR FORMS OF IMPACT MAY CAUSE DAMAGE RESULTING IN INCORRECT OPERATION OF THE UNIT. IF YOU SUSPECT DAMAGE TO HAVE OCCURRED, PLEASE PERFORM CHECKS [AS OUTLINED IN THE MANUAL] BEFORE CONNECTION TO A PATIENT".
(B)(4). THE RD900AEU NEOPUFF INFANT RESUSCITATOR IS EN ROUTE TO FISHER & PAYKEL HEALTHCARE FOR INVESTIGATION. WE WILL PROVIDE A FOLLOW-UP REPORT ONCE WE RECEIVE THE COMPLAINT DEVICE AND HAVE COMPLETED OUR INVESTIGATION.
A MEDICAL CENTER IN (B)(6) REPORTED THAT THE MANOMETER OF AN RD900AEU NEOPUFF INFANT RESUSCITATOR WAS STICKING. NO PATIENT CONSEQUENCE WAS REPORTED.
A MEDICAL CENTER IN (B)(6) REPORTED THAT THE MANOMETER OF AN (B)(4) NEOPUFF INFANT RESUSCITATOR WAS STICKING. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NEOPUFF INFANT RESUSCITATOR | BTL | BTL | FISHER & PAYKEL HEALTHCARE LIMITED | RD900AEU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |