FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2070045 · Received April 26, 2011

Report

Report Number
1423500-2011-05058
Event Type
Malfunction
Date Received
April 26, 2011
Date of Event
April 4, 2011
Report Date
April 4, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PER THE CUSTOMER THE SAMPLE WAS DISCARDED AND THE LOT NUMBER WAS UNKNOWN, THEREFORE NO EVALUATION OR BATCH REVIEW COULD BE PERFORMED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). CUSTOMER REPORTED A LOW DRAIN VOLUME ALARM. PER THE CALL SCRIPT THE ASSIGNABLE CAUSE WAS DETERMINED TO BE AIR IN PATIENT LINE. AS A SAMPLE WAS NOT AVAILABLE FOR EVALUATION AND THE PATIENT DID NOT DESCRIBE ANY TYPE OF USE ERROR THAT MIGHT HAVE CAUSED THE ALARM, THE COMPLAINT COULD NOT BE CONFIRMED. THE ROOT CAUSE WAS UNDETERMINED. THE LOT NUMBER IS UNKNOWN; THEREFORE A BATCH REVIEW CANNOT BE PERFORMED. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Description of Event or Problem · 1

DURING TROUBLESHOOTING OF A LOW DRAIN VOLUME (LDV) ALARM THAT APPEARED ON THE HOMECHOICE (HC) DISPLAY DURING INITIAL DRAIN, THE HOME PATIENT (HP) REPORTED THAT THERE WAS AIR IN PATIENT LINE. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED THE HP IN ENDING THERAPY. THE HP TO RESTART WITH NEW SUPPLIES. THERE WAS PATIENT INVOLVEMENT BUT THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL REPORT. DURING A FOLLOW-UP WITH THE HP ON (B)(6) 2011 REGARDING THE REPORTED EVENT, IT WAS REVEALED THAT SHE WAS FINE AFTER AND STARTING OVER AGAIN. THE HP DID CONTACT HER NURSE. THE HP STATED SHE DID NOT HAVE ANY INJURY OR NEED ANY MEDICAL INTERVENTION DUE TO THIS EVENT. SHE STATED THAT SHE HAS BEEN DOING FINE WITH THERAPY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 33 YR HOMECHOICE CYCLER