ARCHITECT I2000 ANALYZER
Report
- Report Number
- 3005094123-2011-00004
- Event Type
- Malfunction
- Date Received
- April 26, 2011
- Date of Event
- March 22, 2011
- Report Date
- April 4, 2011
- Manufacturer
- ABBOTT MANUFACTURING, INC.
- Product Code
- DHA
- PMA / PMN Number
- K983212
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). EVALUATION: WASH ZONE MANIFOLDS, WASH ZONE 2 PROBES. A FIELD SERVICE REPRESENTATIVE (FSR) WAS DISPATCHED TO THE ACCOUNT. THE FSR REPLACED THE WASH ZONE MANIFOLDS, REPLACED THE PROBES IN WASH ZONE 2 AND CLEANED THE PROCESS PATH AFTER WHICH NO FURTHER ISSUES WERE REPORTED. CUSTOMER COMPLAINT DATA WAS REVIEWED AND NO ADVERSE TRENDS WERE IDENTIFIED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT DEFICIENCY.
(B)(4). AN INVESTIGATION IS IN PROCESS. A FOLLOWUP REPORT WILL BE SUBMITTED WHEN THE INVESTIGATION IS COMPLETE.
(B)(4). THE INITIAL REPORT WAS SUBMITTED UNDER BRAND NAME ABBOTT ARCHITECT TOTAL B-HCG REAGENT, (B)(4) MANUFACTURING SITE. IT WAS DETERMINED THAT THE INVESTIGATION SHOULD BE PERFORMED ON BRAND NAME ABBOTT ARCHITECT I2000 ANALYZER, (B)(4) MANUFACTURING SITE.
THE ACCOUNT STATED THAT AN INITIAL ARCHITECT TOTAL B-HCG RESULT OF 69.82 MIU/ML WAS GENERATED. THE SAMPLE WAS REPEATED AND A RESULT OF < 1.2 MIU/ML WAS GENERATED. THERE WAS NO REPORT OF IMPACT TO PATIENT MANAGEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ARCHITECT I2000 ANALYZER | AUTOMATED IMMUNOASSAY ANALYZER | DHA | ABBOTT MANUFACTURING, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | ARCHITECT TOTAL B-HCG 7K78-25 LOT UNKOWN| ARCHITECT TOTAL B-HCG 7K78-25 LOT UNKOWN| ARCHITECT I2000 LN 3M74-01 SN (B)(4) |