FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® EDTA 2K

MDR report key: 20695405 · Received November 15, 2024

Report

Report Number
1917413-2024-01065
Event Type
Malfunction
Date Received
November 15, 2024
Date of Event
October 18, 2024
Report Date
December 6, 2024
Manufacturer
BECTON, DICKINSON & CO. (BROKEN BOW)
Product Code
JKA
UDI-DI
00382903678464
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

E1: INITIAL REPORTER FACILITY NAME: (B)(6). A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Additional Manufacturer Narrative · 0

THE FOLLOWING FIELDS WERE UPDATED DUE TO ADDITIONAL INFORMATION: D9: DEVICE AVAILABLE FOR EVALUATION: YES. H.3 DEVICE EVAL BY MANUFACTURER? YES. D9: RETURNED TO MANUFACTURER ON: 14-NOV-2024. INVESTIGATION SUMMARY: BD RECEIVED 1 SAMPLE AND EVALUATED 4 PHOTOS FOR INVESTIGATION. THE PHOTOS WERE REVIEWED AND THE CUSTOMER¿S INDICATED FAILURE MODE FOR EMBEDDED FOREIGN MATTER WAS OBSERVED. ADDITIONALLY, THE CUSTOMER SAMPLES WERE EVALUATED BY VISUAL EXAMINATION AND THE INDICATED FAILURE MODE FOR EMBEDDED FOREIGN MATTER WITH THE INCIDENT LOT WAS OBSERVED. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT HAS BEEN CONFIRMED FOR THE INDICATED FAILURE MODE EMBEDDED FOREIGN MATTER. BD WAS NOT ABLE TO IDENTIFY AN EXACT ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED PRIOR TO USING BD VACUTAINER® EDTA 2K A BLACK FOREIGN MATTER WAS FOUND IN 1 TUBE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Description of Event or Problem · 0

IT WAS REPORTED PRIOR TO USING BD VACUTAINER® EDTA 2K A BLACK FOREIGN MATTER WAS FOUND IN 1 TUBE. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
91854 BD VACUTAINER® EDTA 2K BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO. (BROKEN BOW) 4045118 00382903678464

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown