- Report Number
- 9616084-2024-00071
- Event Type
- Malfunction
- Date Received
- November 14, 2024
- Date of Event
- October 15, 2024
- Report Date
- November 14, 2024
- Manufacturer
- SUNRISE MEDICAL (US) LLC
- Product Code
- IOR
- PMA / PMN Number
- K072153
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- OTHER
Narratives
BACKGROUND INFORMATION: QUICKIE QX/QXI WHEELCHAIR OWNER'S MANUAL, REV. H, PAGE 14 STATES: "INSPECT WHEEL LOCKS WEEKLY PER THE MAINTENANCE CHART. DO NOT USE YOUR CHAIR UNLESS YOU ARE SURE BOTH WHEEL LOCKS CAN FULLY ENGAGE. A WHEEL LOCK THAT IS NOT CORRECTLY ADJUSTED MAY ALLOW YOUR CHAIR TO ROLL OR TURN UNEXPECTEDLY. WHEEL LOCKS MUST BE ADJUSTED AFTER MAKING SURE THE TIRES HAVE THE CORRECT AIR PRESSURE. WHEN FULLY ENGAGED, THE ARM SHOULD BE EMBEDDED INTO THE TIRE AT LEAST 1/8" TO BE EFFECTIVE" AND "IF YOU FIND THE WHEEL LOCKS HAVE SLIPPED OR ARE NOT WORKING CORRECTLY CONTACT YOUR SUNRISE MEDICAL AUTHORIZED DEALER FOR PROPER ADJUSTMENT." QUICKIE QX/QXI WHEELCHAIR OWNER'S MANUAL, REV. H, PAGE 28 STATES: "WE WARRANT ALL SUNRISE-MADE PARTS AND COMPONENTS OF THIS WHEELCHAIR AGAINST DEFECTS IN MATERIALS AND WORKMANSHIP FOR ONE YEAR FROM THE DATE OF FIRST CONSUMER PURCHASE." DISCUSSION: IN EVALUATING THE COMPLAINT, THE TECHNICIAN REPORTED THE LEFT-SIDE WHEEL LOCK'S NUT WAS MISSING FROM THE ASSEMBLY. THE TECHNICIAN STATED THE WHEEL LOCK WAS HANGING FROM THE WHEELCHAIR AND WAS UNSURE WHETHER THE HARDWARE WAS STRIPPED. IT WAS DETERMINED THAT BASED ON SIMILAR COMPLAINTS AND PRODUCTS RECEIVED, THE POTENTIAL CAUSE COULD BE HARDWARE LOOSENING OVER TIME WHICH COULD LEAD TO INSUFFICIENT WHEEL LOCKING FORCE. A DHR REVIEW FOR THIS PRODUCT WAS CONDUCTED AND NO ABNORMALITIES, DEVIATIONS OR NCMR'S RELATED TO THE CLAIM WERE IDENTIFIED IN THE MANUFACTURING PROCESS. BASED ON THE OWNER'S MANUAL, THE USER OR DEALER SHOULD BE PERFORMING MAINTENANCE TO REDUCE THE RISK OF WHEEL LOCKS BECOMING NON-FUNCTIONAL OR BROKEN. THE AGE OF THE CHAIR AT THE TIME OF THIS COMPLAINT WAS APPROXIMATELY ONE HUNDRED AND TWO (102) DAYS. ACCORDING TO THE QUICKIE QXI WHEELCHAIR OWNER'S MANUAL, SUNRISE MEDICAL PROVIDES A ONE-YEAR WARRANTY FROM DATE OF PURCHASE FOR ALL SUNRISE-MADE PARTS AND COMPONENTS THAT HAVE DEFECTS IN THE MATERIAL OR WORKMANSHIP. THE DEALER REQUESTED A WARRANTY REPLACEMENT OF THE LEFT-SIDE WHEEL LOCK ASSEMBLY. THERE WERE NO INJURIES OR ADVERSE IMPACT TO THE USER REPORTED. CONCLUSION: THE LOOSENING OF THE WHEEL LOCK HARDWARE OVER TIME IS THE PRIMARY POTENTIAL CAUSE IDENTIFIED. THE PRODUCT IN QUESTION MET ALL PRODUCT SPECIFICATIONS BEFORE RELEASE FOR DISTRIBUTION AT THE TIME OF SHIPPING TO THE DEALER. THIS DEVICE IS USED FOR TREATMENT, NOT DIAGNOSIS. THERE IS NO CLAIM OF INJURY IN THIS CASE. BECAUSE THE FAILURE MODE OF NON-FUNCTIONING WHEEL LOCKS HAS BEEN PREVIOUSLY REPORTED, PER 21 CFR 803.50, THIS MDR IS BEING FILED.
THE TECHNICIAN REPORTED THE LEFT-SIDE WHEEL LOCK'S NUT WAS MISSING FROM THE ASSEMBLY. THE TECHNICIAN STATED THE WHEEL LOCK WAS HANGING FROM THE WHEELCHAIR AND WAS UNSURE WHETHER THE HARDWARE WAS STRIPPED. THE DEALER REQUESTED A WARRANTY REPLACEMENT OF THE LEFT-SIDE WHEEL LOCK ASSEMBLY. NO INJURIES OR ADVERSE IMPACT TO THE USER WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 2285435 | QUICKIE | MECHANICAL WHEELCHAIR | IOR | SUNRISE MEDICAL (US) LLC | QUICKIE QXI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |