FDA Adverse Event Malfunction Summary report: N

SPIRAL COMBINATION HAMMER 500 GRAMS

MDR report key: 20677242 · Received November 13, 2024

Report

Report Number
8030965-2024-13869
Event Type
Malfunction
Date Received
November 13, 2024
Date of Event
November 4, 2024
Manufacturer
SYNTHES GMBH
Product Code
FZY
UDI-DI
07611819499171
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H3, H4, H6: THE PRODUCT WAS NOT RETURNED TO MEDTECH ORTHOPAEDICS, HOWEVER PHOTOS WERE PROVIDED FOR REVIEW. E NOVEDAD OF 885139. THE PHOTO INVESTIGATION REVEALED THAT '( 03.010.522, COMBINED HAMMER 500G) HAD BROKEN. THE OBSERVED CONDITION OF THE DEVICE WAS CONSIST WITH A RANDOM COMPONENT FAILURE THAT MAY HAVE BEEN CAUSED BY EXPOSURE TO UNINTENDED FORCE. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS (CONFIRMED) AS THE OBSERVED CONDITION OF THE (COMBINED HAMMER 500G) WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, POTENTIAL CAUSE CAN BE ATTRIBUTED TO CAUSE TRACE TO COMPONENT FAILURE AND IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY REVIEW : PART # 03.010.522, LOT # 9692937, MANUFACTURING SITE: WERK VILLMERGEN LOGISTIK, RELEASE TO WAREHOUSE DATE: 22.OCT.2015, EXPIRATION DATE: N/A, SUPPLIER: (B)(4). A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED ARTICLE LOT AND NO NON-CONFORMANCES WERE IDENTIFIED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING A PROCEDURE THE HAMMER SPLIT WHILE BEING USED. NO DISCOMFORT OF THE SPECIALIST. A DIFFERENT HAMMER WAS USED. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED STATES THAT THE SURGERY WAS COMPLETED SUCCESSFULLY WITH NO SURGICAL DELAY. THERE WERE 2 LARGE FRAGMENTS LEFT, WHICH DO NOT REMAIN INSIDE THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2432092 SPIRAL COMBINATION HAMMER 500 GRAMS SURGICAL HAMMER FZY SYNTHES GMBH 9692937 07611819499171

Patients

Seq Age Sex Outcome Treatment
1 85 YR Male