FDA Adverse Event Malfunction Summary report: N

UNICEL® DXI 800 ACCESS® IMMUNOASSAY SYSTEM

MDR report key: 2067042 · Received April 22, 2011

Report

Report Number
2122870-2011-01028
Event Type
Malfunction
Date Received
April 22, 2011
Date of Event
March 24, 2011
Report Date
March 24, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K023764
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

AFP QC WAS WITHIN THE CUSTOMER'S ESTABLISHED RANGES PRIOR TO THE EVENT AND OUT OF RANGE HIGH AFTER THE EVENT. SERVICE WAS DISPATCHED AND FOUND THAT THE DILUTION CORRECTION FACTORS FOR DIL-AFP WERE FOUND TO BE TEN TIMES GREATER THAN THE ACCEPTABLE RANGE. A DEFINITIVE ROOT CAUSE HAS NOT BEEN DETERMINED TO DATE FOR THIS EVENT AND IS STILL BEING INVESTIGATED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC., (BCI) REGARDING A HIGHER THAN EXPECTED ALPHA-FETOPROTEIN (AFP) RESULT GENERATED BY THE UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM FOR ONE PATIENT. SUBSEQUENT TESTING ON AN ALTERNATE INSTRUMENT PRODUCED A LOWER RESULT THAT MATCHED THE PATIENT'S CLINICAL PRESENTATION. THE HIGH RESULT WAS NOT REPORTED OUT OF THE LAB. THE CUSTOMER DID NOT RECEIVE ANY REPORT OF PATIENT INJURY REQUIRING MEDICAL INTERVENTION OR CHANGE TO PATIENT TREATMENT ATTRIBUTED OR CONNECTED TO THIS EVENT.