FDA Adverse Event Malfunction Summary report: N

BD INSYTE AUTOG BC

MDR report key: 20669285 · Received November 12, 2024

Report

Report Number
1710034-2024-01328
Event Type
Malfunction
Date Received
November 12, 2024
Date of Event
October 25, 2024
Report Date
February 3, 2025
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903825349
PMA / PMN Number
K201075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
501

Narratives

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. CUSTOMER ALSO PROVIDE LOT NUMBER 4190991 WHICH DOES NOT PULL A VALID MATERIAL NUMBER FOR THIS DEVICE.

Additional Manufacturer Narrative · 0

INVESTIGATION RESULTS: OUR QUALITY ENGINEER INSPECTED THE REPRESENTATIVE SAMPLES SUBMITTED FOR EVALUATION. BD RECEIVED TWO SEALED 20GA X 1.16IN. INSYTE AUTOGUARD BC UNITS FROM LOT NUMBER 4260454. A VISUAL INSPECTION DID NOT DISCOVER ANY DEFECTS TO THE RETURNED UNITS THAT RESEMBLED OR DISPLAYED THE NEEDLE PIERCING THROUGH THE CATHETER WALL. SHOULD YOU AGAIN EXPERIENCE ANY PROBLEMS WITH OUR PRODUCT WE WOULD APPRECIATE THE OPPORTUNITY TO CONDUCT A THOROUGH ANALYSIS. EXAMINATION OF THE ACTUAL PRODUCT INVOLVED MAY PROVIDE CLARIFICATION AS TO THE CAUSE FOR THE REPORTED FAILURE. A DEVICE HISTORY RECORD REVIEW SHOWED NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING THE PRODUCTION OF THIS BATCH.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD NSYTE AUTOG BC NEEDLE PIERCING THE CATHETER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE IV CATHETER IS SHEERING THROUGH NEEDLE. 29OCT: THE SELF-OCCLUDING IV CATHETERS ARE SHEERING THE INSIDE OF THE PLASTIC CATHETER. THIS WAS NOTICED DURING USE. NO PATIENT HARM.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1058892 BD INSYTE AUTOG BC PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 4260454 00382903825349

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown