FDA Adverse Event Malfunction Summary report: N

CORE 500 DIGITAL STETHOSCOPE

MDR report key: 20665385 · Received November 12, 2024

Report

Report Number
3011651230-2024-00001
Event Type
Malfunction
Date Received
November 12, 2024
Report Date
October 14, 2024
Manufacturer
EKO HEALTH, INC.
Product Code
DQD
UDI-DI
00850010298160
PMA / PMN Number
K230111
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE REFERENCED DEVICE WAS FORWARDED TO AN INDEPENDENT LABORATORY FOR FURTHER INVESTIGATION. THE INVESTIGATION FOUND A SIGNIFICANT AMOUNT OF CHARRED FOREIGN MATERIALS WHICH HAVE BEEN IDENTIFIED AS CHLORINE, PHOSPHORUS, SODIUM, AND FLUORINE INSIDE THE PORT OF THE USB-C CONNECTOR. THESE FOREIGN MATERIALS WERE CONSISTENT WITH HOUSEHOLD CLEANING SOLUTIONS. VISUAL INSPECTION OF THE REFERENCED DEVICE FOUND BURN MARKS AROUND THE SHIELD. THE USB-C PORT DID NOT EXHIBIT ANY SIGNS OF OVERHEATING. HOWEVER, SOME BROWNISH RESIDUES WERE OBSERVED. NO SIGNS OR PIN DAMAGE OR ANOMALOUS HIGH DENSITY RESIDUES WERE OBSERVED ON THE PINS WITHIN THE USB-C PORT. THE REPORTED INCIDENT WAS MOST LIKELY CAUSED BY THE FOREIGN MATERIALS AS THEY WERE OBSERVED IN SEVERAL LOCATIONS WITH THE USB-C CONNECTOR WHICH MIGHT HAVE CONTRIBUTED TO ELECTROCHEMICAL METAL MIGRATION AND CAUSE FAULT BETWEEN THE POWER AND GND PINS. BASED ON THE INVESTIGATION FINDINGS, IT WAS DETERMINED THE REPORTED INCIDENT WAS ATTRIBUTED TO INADVERTENT CONTAMINATION OF THE USB-C PORT.

Additional Manufacturer Narrative · 0

CUSTOMER ITEM RETURNED WITH EKO SUPPLIED USB-C CHARGING CABLE AND A THIRD-PARTY LAX BRAND CHARGER ON (B)(6)2024. A PRELIMINARY INITIAL VISUAL EXAMINATION REVEALED THAT THE USB-C CABLE EXPERIENCED A THERMAL EVENT AND DEMONSTRATED TRACES OF MELTED PLASTIC. THE CORE 500 CHESTPIECE DID NOT EXHIBIT SIGNS OF BURNING/CHARRING/MELTING, AND THE CHESTPIECE APPEARS TO FUNCTION NORMALLY WITH A KNOWN-GOOD CHARGING CABLE. ALL THREE ITEMS HAVE BEEN FORWARDED TO AN EXTERNAL TEST LABORATORY (XPONENT) FOR SPECIALIZED ELECTRICAL COMPONENT EXAMINATION ((B)(6)2024). A SUPPLEMENTARY REPORT WILL BE SUBMITTED AT A LATER DATE ONCE THE INVESTIGATION RESULT BECOMES AVAILABLE.

Description of Event or Problem · 0

CUSTOMER REPORTED THAT THE REFERENCED DEVICE OVERHEATED AND THE USB-C CHARGING CABLE MELTED DURING CHARGING. NO DEATH OR SERIOUS INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1658785 CORE 500 DIGITAL STETHOSCOPE CORE 500 DIGITAL STETHOSCOPE DQD EKO HEALTH, INC. COR500 00850010298160

Patients

Seq Age Sex Outcome Treatment
1 NA Female Other