BD INSYTE AUTOG BC
Report
- Report Number
- 1710034-2024-01309
- Event Type
- Malfunction
- Date Received
- November 12, 2024
- Date of Event
- October 23, 2024
- Report Date
- December 10, 2024
- Manufacturer
- BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
- Product Code
- FOZ
- UDI-DI
- 00382903825233
- PMA / PMN Number
- K201075
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. E1. ADDRESS INFORMATION WAS NOT PROVIDED, THEREFORE, XX WAS USED AS A PLACE HOLDER.
NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 382523 AND LOT NUMBER 4102965. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. BASED ON THE AVAILABLE INFORMATION, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.
IT WAS REPORTED THAT BD INSYTE AUTOG BC LEAKED AT THE HUB/IV TUBING JUNCTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: TODAY WE HAD IN INCIDENT WITH ONE OF THE BD INSYTE AUTOGUARD 22 GAUGE IV CATHETERS. MY ASU RN INSERTED THE CATHETER, WAS ABLE TO OBTAIN A FLASH, SO TOOK THE NEEDLE OUT AND WHEN SHE TRIED TO HOOK UP THE IVF, IT WAS LEAKING AT THE HUB. SHE UNSCREWED THE IVF FROM THE HUB OF THE CATHETER, AND TRIED TO SCREW IN AGAIN, AND IT STILL LEAKED. ULTIMATELY, SHE HAD TO REMOVE THE CATHETER COMPLETELY AND INSERT ANOTHER ONE, AT ANOTHER SITE.
NO ADDITIONAL INFORMATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1059611 | BD INSYTE AUTOG BC | PERIPHERAL IV CATHETERS | FOZ | BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. | 4102965 | 00382903825233 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |