FDA Adverse Event Malfunction Summary report: N

BD INSYTE AUTOG BC

MDR report key: 20664792 · Received November 12, 2024

Report

Report Number
1710034-2024-01309
Event Type
Malfunction
Date Received
November 12, 2024
Date of Event
October 23, 2024
Report Date
December 10, 2024
Manufacturer
BECTON DICKINSON INFUSION THERAPY SYSTEMS INC.
Product Code
FOZ
UDI-DI
00382903825233
PMA / PMN Number
K201075
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ME, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H.3. A FOLLOW UP MDR WILL BE SUBMITTED IF ADDITIONAL INFORMATION, A DEVICE EVALUATION, OR A DEVICE HISTORY REVIEW IS COMPLETED. E1. ADDRESS INFORMATION WAS NOT PROVIDED, THEREFORE, XX WAS USED AS A PLACE HOLDER.

Additional Manufacturer Narrative · 0

NO PHOTOS OR PHYSICAL SAMPLES THAT DISPLAY THE REPORTED CONDITION WERE AVAILABLE FOR INVESTIGATION. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED BY OUR QUALITY ENGINEER TEAM FOR PROVIDED MATERIAL NUMBER 382523 AND LOT NUMBER 4102965. THE REVIEW DID NOT REVEAL ANY DETECTED ABNORMALITIES DURING THE PRODUCTION PROCESS THAT COULD HAVE CONTRIBUTED TO THIS DEFECT AND ALL QUALITY TESTS WERE FOUND TO BE WITHIN SPECIFICATION. BASED ON THE AVAILABLE INFORMATION, AN EXACT CAUSE FOR THIS INCIDENT COULD NOT BE IDENTIFIED. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR QUALITY TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD INSYTE AUTOG BC LEAKED AT THE HUB/IV TUBING JUNCTION. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: TODAY WE HAD IN INCIDENT WITH ONE OF THE BD INSYTE AUTOGUARD 22 GAUGE IV CATHETERS. MY ASU RN INSERTED THE CATHETER, WAS ABLE TO OBTAIN A FLASH, SO TOOK THE NEEDLE OUT AND WHEN SHE TRIED TO HOOK UP THE IVF, IT WAS LEAKING AT THE HUB. SHE UNSCREWED THE IVF FROM THE HUB OF THE CATHETER, AND TRIED TO SCREW IN AGAIN, AND IT STILL LEAKED. ULTIMATELY, SHE HAD TO REMOVE THE CATHETER COMPLETELY AND INSERT ANOTHER ONE, AT ANOTHER SITE.

Description of Event or Problem · 0

NO ADDITIONAL INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1059611 BD INSYTE AUTOG BC PERIPHERAL IV CATHETERS FOZ BECTON DICKINSON INFUSION THERAPY SYSTEMS INC. 4102965 00382903825233

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown