FDA Adverse Event Malfunction Summary report: N

AUTOMATED IMPELLA CONTROLLER

MDR report key: 20654157 · Received November 11, 2024

Report

Report Number
1220648-2024-22896
Event Type
Malfunction
Date Received
November 11, 2024
Date of Event
May 27, 2022
Report Date
November 11, 2024
Manufacturer
ABIOMED, INC.
Product Code
OZD
UDI-DI
00813502010022
PMA / PMN Number
P140003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE INVESTIGATION FOR THE REPORTED ISSUE HAS BEEN COMPLETED. THE PRODUCT WAS RETURNED, AND THE ISSUE WAS NOT CONFIRMED. CONSOLE LOGS FROM THE REPORTED DAY OF EVENT ARE CONSISTENT WITH COMPLAINT AND SHOW CONTROLLER ERROR ALARM (#24) DUE TO KBD COMM LOSS, UPON CONSOLE BOOT-UP. THIS IS PRECEDED BY COMMUNICATION ERROR (IMC-KBD ERROR: 0XA4 0X01 0X00), AND THE FACT THAT CONSOLE¿S BOOT-UP COUNT WAS NOT CONSECUTIVE, INDICATING PRIOR FAILED BOOT-UP SEQUENCE: AFTER CONSOLE WAS MANUALLY POWER-CYCLED, THE ISSUE DID NOT RE-OCCUR. CONSOLE WAS TESTED IN DANVERS, AND POWER-CYCLED SEVERAL TIMES, BUT THE ISSUE WAS NOT REPRODUCED. HOWEVER, THIS IS A KNOWN RARE PATTERN FAILURE WHICH RESULTS FROM MOMENTARY COMMUNICATION BREAKDOWN DURING INITIALIZATION OF KBD TO 4-IN-1 COMMUNICATION. THIS ISSUE COULD BE CAUSED BY EITHER OF THE COMMUNICATING DEVICES/MICROPROCESSORS: MSC1210 ON THE 4-IN-1 OR AT42QT1060 ON THE KBD ASSEMBLY. A COMPROMISED RIBBON CABLE BETWEEN THESE TWO DEVICES IS ALSO A PLAUSIBLE CAUSE. THE CAUSE OF THE ISSUE WAS NOT DETERMINED SINCE ISSUE WAS UNABLE TO BE REPRODUCED. THIS REPORT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF HISTORICAL RECORDS.

Description of Event or Problem · 0

THE USER FACILITY REPORTED FOR AN AUTOMATED IMPELLA CONTROLLER (AIC) THAT THERE WAS A CONTROLLER ERROR ALARM DURING PREPARATION OF THE DEVICE. A NEW CONSOLE WAS USED WITH NO ADVERSE IMPACT TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1721414 AUTOMATED IMPELLA CONTROLLER CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS OZD ABIOMED, INC. AUTOMATED IMPELLA CONTROLLER 1200770 00813502010022

Patients

Seq Age Sex Outcome Treatment
1 38 YR Male