FDA Adverse Event
Malfunction
Summary report: N
DARBY
MDR report key: 20651245
·
Received November 11, 2024
Report
- Report Number
- 3003444492-2024-00003
- Event Type
- Malfunction
- Date Received
- November 11, 2024
- Date of Event
- September 17, 2024
- Report Date
- October 24, 2024
- Manufacturer
- PRIMA DENTAL GROUP
- Product Code
- EJL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- DENTIST
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
A SAMPLE OF RETURNED DEVICES WERE TESTED FOR NECK STRENGTH AND ALL DEVICES PASSED THE REQUIRED SPECIFICATION.
Description of Event or Problem · 0
DARBY - 9525118/010PK CARBIDE BURS FG 4P - LOT 1182470 - BURS ARE BEING RETURNED FROM A CUSTOMER BECAUSE THE BURS ARE BREAKING. RETURN OF 59 PACKS OF 10 BURS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 944819 | DARBY | FG 4P - 9525118 | EJL | PRIMA DENTAL GROUP | CPEARS014FG170D | 1182470 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 123 YR | Unknown |