SPRINT FIDELIS
Report
- Report Number
- 2649622-2011-07234
- Event Type
- Death
- Date Received
- April 20, 2011
- Date of Event
- November 20, 2010
- Report Date
- March 15, 2024
- Manufacturer
- MEDTRONIC PUERTO RICO, INC.
- Product Code
- LWS
- PMA / PMN Number
- ASKU
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- HEALTH PROFESSIONAL
- Health Professional
- Yes
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY FOR (B)(4): THE ACTUAL DEVICE WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. POWER ON RESET (POR) - POWER ON RESET PARAMETERS; CRITICAL RAM PARITY ERROR CAUSED POWER ON RESET ON (B)(6) 2010 AT ADDRESS 1A EA. POR SEVERITY: LOW. THE DEVICE SHOULD BE ABLE TO FULLY RECOVER AFTER THE RESET. THE DEVICE WAS NOT RETURNED, BUT DIAGNOSTIC INFORMATION IS CONSISTENT WITH REPORTED EVENT.
IT WAS REPORTED THAT THERE WAS AN ELECTRICAL RESET. LATER REVIEW OF MANUFACTURER'S DATABASE REVEALED THE PATIENT DIED APPROXIMATELY FIVE WEEKS AFTER THIS EVENT. THE CAUSE OF DEATH HAS BEEN REQUESTED AND NOT RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SPRINT FIDELIS | IMPLANTABLE TACHY LEAD | LWS | MEDTRONIC PUERTO RICO, INC. | 6949 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 85 YR | Male | Death | 6949 IMPLANTABLE TACHY LEAD| 6949 IMPLANTABLE TACHY LEAD |