TPRLC 133 MP TYPE1 PPS HO 12.0
Report
- Report Number
- 0001825034-2024-02631
- Event Type
- Malfunction
- Date Received
- November 8, 2024
- Date of Event
- October 15, 2024
- Report Date
- March 4, 2025
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- KWA
- UDI-DI
- 00880304513471
- PMA / PMN Number
- K110400
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED/CORRECTED UPDATED: B4 G3 G6 H2 H6 H11 VISUAL EVALUATION OF THE RETURNED PRODUCT/PROVIDED PHOTOS IDENTIFIED DAMAGE TO THE STERILE PACKAGING (BLISTER). STERILITY HAS NOT BEEN COMPROMISED. REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING RELATED TO THE REPORTED EVENT. THE REPORTED EVENT DID NOT OCCUR IN AN OPERATING ROOM OR AS PART OF A MEDICAL PROCEDURE; MEDICAL RECORDS ARE NOT AVAILABLE FOR REVIEW. THE CONDITION OF THE DEVICE WHEN IT LEFT ZIMMER BIOMET IS CONSIDERED CONFORMING TO SPECIFICATION. THE ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO TRANSIT DAMAGE AND A PACKAGING DESIGN ISSUE. THE REPORTED EVENT HAS BEEN CONFIRMED BY EVALUATION OF THE RETURNED PRODUCT AND PROVIDED PHOTOS. THE DEVICE EVALUATION FOUND THE PRODUCT TO BE CONFORMING AND STERILITY NOT COMPROMISED. FURTHER, THE EVENT DID NOT CONTRIBUTE TO A SERIOUS INJURY; THEREFORE, THIS WOULD NOT BE CONSIDERED A REPORTABLE EVENT. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
(B)(4). D10: 51-106140 ITEM NAME TPRLC 133 MP TYPE1 PPS SO 14.0 LOT # 6994250. 51-103150 ITEM NAME TPRLC 133 T1 PPS SO 15X150MM LOT # 7005059. 51-145120 ITEM NAME TPRLC XR MP T1 PPS 12X109MM LOT# 7320778. 51-103140 ITEM NAME TPRLC 133 T1 PPS SO 14X148MM LOT # 6423079. G2: FOREIGN: JAPAN. CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
THERE IS NO UPDATE TO THE PRIOR EVENT DESCRIPTION PROVIDED.
IT WAS REPORTED WHEN INSPECTING CIRCULATED PRODUCT THAT THE STERILE PACKAGING WAS FOUND DAMAGED. THERE WAS NO PATIENT INVOLVEMENT. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1951833 | TPRLC 133 MP TYPE1 PPS HO 12.0 | PROSTHETIC, HIP | KWA | ZIMMER BIOMET, INC. | N/A | 7096392 | 00880304513471 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Prefer Not To Disclose |