FDA Adverse Event Injury Summary report: N

ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO

MDR report key: 2063872 · Received April 15, 2011

Report

Report Number
2953200-2011-00814
Event Type
Injury
Date Received
April 15, 2011
Date of Event
March 16, 2011
Report Date
March 16, 2011
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P990020
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION, RESULTS: OCCLUSION. MILDLY TORTUOUS AND MILDLY CALCIFIED, AORTIC NECK ANGULATION, CALCIFICATION. AORTIC CUFF WAS IMPLANTED LOWER THAN INTENDED AND OCCLUDED THE CONTRALATERAL LIMB. CONCLUSIONS: MILDLY TORTUOUS AND MILDLY CALCIFIED, AORTIC NECK ANGULATION, CALCIFICATION. AORTIC CUFF WAS IMPLANTED LOWER THAN INTENDED AND OCCLUDED THE CONTRALATERAL LIMB.

Description of Event or Problem · 1

AN ANEURX STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM. VESSEL MORPHOLOGY WAS REPORTED AS MILDLY TORTUOUS AND MILDLY CALCIFIED, WITH AN AORTIC NECK, CALCIFICATION AND ANGULATION SLIGHTLY LESS THAN 45 DEGREES. THE AORTIC NECK DIAMETER WAS 21-23 MM WITH A LENGTH OF 20 MM. A PROXIMAL TYPE I ENDOLEAK WAS NOTED AND A 26 CM CUFF WAS PLACED WHICH CORRECTED THE ENDOLEAK (REF MFR # 2953200-2011-00125). IT WAS REPORTED THE PATIENT HAD A ROUTINE CT THREE MONTHS POST-IMPLANT, WHICH DEMONSTRATED A PROXIMAL TYPE I ENDOLEAK AND THERE WAS OCCLUSION OF THE CONTRALATERAL LIMB. (REF MFR # 2953200-2011-00813). IT APPEARS THAT THE AORTIC CUFF THAT WAS IMPLANTED FOR THE REPORTED INITIAL TYPE I ENDOLEAK WAS IMPLANTED LOWER THAN INTENDED OCCLUDING OFF THE CONTRALATERAL LIMB. THE PATIENT WAS BACK AT A LATER DATE AND AN ENDURANT AORTIC CUFF WAS IMPLANTED AND SUCCESSFULLY RESOLVED THE PROXIMAL ENDOLEAK. CURRENTLY THE PATIENT HAS COLLATERAL CIRCULATION AND NO SYMPTOMS OR ISSUES DUE TO THE OCCLUDED STENT GRAFT. FILMS WERE RECEIVED AND THEIR EVALUATION WAS COMPLETED. REVIEW OF THE THREE MONTH POST-IMPLANT FILMS SHOW CALCIFICATION WITHIN THE AORTIC NECK WHICH MAY HAVE CONTRIBUTED TO THE PROXIMAL TYPE I ENDOLEAK. THE FILMS ALSO CONFIRMED THAT THE CUFF HAD BEEN PLACED LOW IN THE BIFURCATE AORTIC BODY AND WAS COVERING THE CONTRALATERAL LIMB. THE CONTRA LIMB WAS OCCLUDED DISTALLY FROM THE FLOW DIVIDER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANEURX AAADVANTAGE STENT GRAFT SYSTEM - HYDRO MIH MEDTRONIC IRELAND NA V00553178

Patients

Seq Age Sex Outcome Treatment
1 85 YR Required Intervention