ACUSON REDWOOD
Report
- Report Number
- 3023245-2024-00062
- Event Type
- Malfunction
- Date Received
- November 8, 2024
- Date of Event
- November 5, 2024
- Report Date
- November 7, 2024
- Manufacturer
- SIEMENS HEALTHINEERS
- Product Code
- IYN
- UDI-DI
- 04056869251264560070
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SN
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
THE INVESTIGATION WAS CONDUCTED WITH PROVIDED INFORMATION AND THE REPORTED ISSUE WAS CAUSED BY A SOFTWARE CODING ERROR OF THE MEASUREMENT RATIO TOOL. ASSESSMENT OF THE SOFTWARE CODE DETERMINED THE RATIO WAS CODED ACCORDING TO MEASUREMENT ORDER, RATHER THAN CODED TO THE SPECIFIED DISTANCE VALUE. PER THE CODE, THE FIRST PERFORMED DISTANCE MEASUREMENT IS ALWAYS DIVIDED BY THE SECOND DISTANCE MEASUREMENT. THEREFORE, WHEN A B-MODE IMAGE DISTANCE IS MEASURED FIRST, AS IN THE CUSTOMER COMPLAINTS, THE INVERSE RATIO VALUE IS CALCULATED AND DISPLAYED. BASED ON THE RESULT OF HRE, THE RISK INDEX 6.5 RESIDUAL RISK IS 2A BUT THE RESULT OF THIS RISK WAS ASSESSED AS 1A. NO CAPA WAS REQUIRED. SEVERITY OF HARM ASSOCIATED WITH THIS IS A LEVEL [1], NEGLIGIBLE INJURY PROBABILITY OF OCCURRENCE OF HARM IS IMPROBABLE [A] RISK INDEX [I] - GENERALLY ACCEPTABLE, BELOW RISK ACCEPTABILITY THRESHOLD THE HRE ASSESSES THE RISK AS ACCEPTABLE FOR NOT PROVIDING A FIELD PROVISION OF A CORRECTIVE SOLUTION SYSTEM, AND THAT CURRENTLY DISTRIBUTED REDWOOD SYSTEMS WILL CONTINUE TO BE ABLE TO MANIFEST THIS DEFECT. THIS ISSUE WAS ADDRESSED IN REDWOOD VA30A. REFERENCE: (B)(4).
AN INVESTIGATION IS ONGOING AT SIEMENS HEALTHINEERS TO IDENTIFY THE ROOT CAUSE OF THE ISSUE. THE REPORT WILL BE SUPPLEMENTED AFTER THE INVESTIGATION IS COMPLETED. REFERENCE# (B)(4).
STRAIN IMAGING EI/B RATIO IS NOT CALCULATED ACCORDING TO MEASUREMENT LABELS. WORKFLOW FOR REPRODUCE: 18L6 BREAST EXAM, VT STRAINING IMAGING. 1. MEASUREMENT MODE, ACTIVATE "EI/B RATIO - D." 2. MEASURE DISTANCE ON B MODE IMAGE D (B) MEASUREMENT VALUE IS DISPLAYED ON IMAGE SCREEN. 3. MEASURE DISTANCE ON ELASTO IMAGE D (E) MEASUREMENT VALUE IS DISPLAYED ON IMAGE SCREEN. 4. THE EI /B DISTANCE RATIO IS CALCULATED, BUT RESULT IS B/EI DISTANCE RATIO, NOT EI/B DISTANCE RATIO.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1877029 | ACUSON REDWOOD | ULTRASOUND DEVICE | IYN | SIEMENS HEALTHINEERS | 11503314 | 04056869251264560070 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |