FDA Adverse Event Injury Summary report: N

ZIMMER M/L TAPER FEMORAL STEM

MDR report key: 2063736 · Received April 14, 2011

Report

Report Number
1822565-2011-00949
Event Type
Injury
Date Received
April 14, 2011
Report Date
March 21, 2011
Manufacturer
ZIMMER, INC.
Product Code
LPH
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL SUMMARY: THE EXACT ORIGIN OF THE PAIN IS UNK. NO DEVICES OR PHOTOS WERE RECEIVED; THEREFORE, THE CONDITION OF THE COMPONENTS IS UNK. SURGICAL NOTES WERE PROVIDED INDICATING THAT LEG LENGTH MAY BE 2-3 MM LONG BUT IT CANNOT BE CONFIRMED THAT THIS CAUSED OR CONTRIBUTED TO THE PAIN. THE SURGEON ALSO NOTED THAT THE IMPLANTS GAINED SOLID FIXATION. X-RAYS WERE NOT PROVIDED; IT CANNOT BE CONFIRMED WHETHER THE COMPONENTS WERE IMPLANTED WITH THE CORRECT FIT AND ORIENTATION AS PER THE SURGICAL TECHNIQUE. PT FACTORS THAT MAY AFFECT THE PERFORMANCE OF THE COMPONENTS SUCH AS BONE QUALITY, HEIGHT/WEIGHT, ACTIVITY LEVEL, TYPE OF ACTIVITY (LOW IMPACT VS. HIGH IMPACT) ARE UNK. THERE IS INSUFFICIENT INFO TO PERFORM A PROBABLE CAUSE ANALYSIS. A DEFINITIVE ROOT CAUSE CANNOT BE DETERMINED WITH THE INFO PROVIDED. EVAL: REVIEW OF THE DEVICE HISTORY RECORDS DID NOT FIND ANY DEVIATIONS OR ANOMALIES. IT IS NOT SUSPECTED THAT THE PRODUCT FAILED TO MEET SPECS. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY EVIDENCE OF PRODUCT CONTRIBUTION TO THE REPORTED PROBLEM. BASED ON THE INVESTIGATION, THE NEED FOR CORRECTIVE ACTION IS NOT INDICATED. SHOULD ADD'L SUBSTANTIVE INFO BE RECEIVED, THE COMPLAINT WILL BE REOPENED. ZIMMER, INC. CONSIDERS THE INVESTIGATION CLOSED.

Description of Event or Problem · 1

IT IS REPORTED THAT THE PT IS EXPERIENCING PAIN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ZIMMER M/L TAPER FEMORAL STEM LPH ZIMMER, INC. 60718275

Patients

Seq Age Sex Outcome Treatment
1 Other