HEATED OXYGEN THERAPY KIT
Report
- Report Number
- 9611451-2011-00260
- Event Type
- Malfunction
- Date Received
- April 20, 2011
- Date of Event
- March 22, 2011
- Report Date
- March 24, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BTT
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE IS EN ROUTE TO THE MANUFACTURER. WE WILL PROVIDE A FINAL REPORT FOLLOWING RECEIPT OF THE DEVICE AND COMPLETION OF OUR INVESTIGATION.
(B)(4). METHOD: THE RETURNED INSPIRATORY LIMB WAS VISUALLY INSPECTED AND PRESSURE TESTED. RESULTS: A HOLE WITH ROUGH EDGES WAS FOUND IN THE MIDDLE OF THE INSPIRATORY LIMB. THE PRESSURE TEST REVEALED THAT THE CIRCUIT WAS OUT OF SPECIFICATION. A LOT CHECK REVEALED NO OTHER COMPLAINTS FOR THIS LOT NUMBER. CONCLUSION: WE ARE UNABLE TO DETERMINE WHAT CAUSED THE OBSERVED DAMAGE TO THE CIRCUIT, BUT IT MOST LIKELY HAPPENED DURING TRANSPORT OR WHILE IN STORAGE OR USE AT THE CUSTOMER FACILITY. ALL CIRCUITS ARE PRESSURE AND FLOW TESTED BEFORE THEY ARE ALLOWED TO LEAVE THE PRODUCTION LINE. THIS IS AN AUTOMATED PROCESS AND THE CIRCUIT WOULD HAVE MET THE REQUIRED SPECIFICATION AT THE TIME OF PRODUCTION.
A HOSPITAL IN (B)(6) REPORTED THAT THE INSPIRATORY LIMB OF AN RT308 HEATED OXYGEN THERAPY KIT HAD A HOLE IN IT. THIS WAS FOUND BEFORE PATIENT USE.
A HOSPITAL IN THE (B)(6) REPORTED THAT THE INSPIRATORY LIMB OF AN RT308 HEATED OXYGEN THERAPY KIT HAD A HOLE IN IT. THIS WAS FOUND BEFORE PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HEATED OXYGEN THERAPY KIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE LTD | RT308 | 100729 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |