FDA Adverse Event Malfunction Summary report: N

ECHELON 60

MDR report key: 2063652 · Received April 19, 2011

Report

Report Number
3005075853-2011-01581
Event Type
Malfunction
Date Received
April 19, 2011
Date of Event
March 24, 2011
Report Date
April 5, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K051002
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). INCOMPLETE/INTERRUPTED FIRING CYCLE. THE EC60 DEVICE WAS RECEIVED FOR ANALYSIS WITH NO VISUAL NON-CONFORMANCES AND WITH A CARTRIDGE RELOAD PRESENT. THE CARTRIDGE RELOAD WAS RECEIVED PARTIALLY FIRED. A PARTIAL FIRE CAN OCCUR IF THE FIRING SEQUENCE IS INTERRUPTED. THE RETURNED CARTRIDGE RELOAD WAS TESTED FOR FUNCTIONALITY BY RESETTING AND RELOADING IT INTO THE DEVICE. THE FUNCTIONAL TEST DEMONSTRATED THAT THE REMAINING STAPLES IN THE CARTRIDGE RELOAD FORMED PROPERLY AND THE INSTRUMENT ACHIEVED ITS COMPLETE 3-STROKE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. IT SHOULD BE NOTED THAT ALL DEVICES ARE INSPECTED 100% FOR STAPLE PRESENCE BY AN AUTOMATED VISION SYSTEM, AND ARE VISUALLY INSPECTED 100% AS A FINAL CHECK. IN ADDITION, AT FINISHED GOODS THE DEVICES ARE VISUALLY INSPECTED BASED ON A SAMPLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE DEVICE FINISHED THE THIRD FIRING. BUT WHEN THE SURGEON OPENED THE DEVICE, IT WAS FOUND THAT THE FRONT 1/2 STAPLES "COULD NOT COME". ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ECHELON 60 STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. UNK NI

Patients

Seq Age Sex Outcome Treatment
1 RELOAD PRODUCT #- ECR60D