UNICEL® DXI 800 ACCESS® IMMUNOASSAY ANALYZER
Report
- Report Number
- 2122870-2011-01023
- Event Type
- Malfunction
- Date Received
- April 19, 2011
- Date of Event
- March 20, 2011
- Report Date
- March 20, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K023764
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
SERVICE WAS ON SITE (B)(4) 2011. THE FIELD SERVICE ENGINEER (FSE) DISCOVERED THAT THE LEAK WAS COMING FROM THE PERISTALTIC PUMP TUBING RUNNING THROUGH THE WASH BUFFER TRANSFER PUMP. THE FSE REPLACED THE WASH BUFFER TRANSFER PUMP TUBING ALONG WITH ALL OTHER TUBING WITHIN THE FLUIDICS DRAWER OF THE DXI 800. THE FSE ALSO REPLACED ALL ASPIRATE PROBES AND THE ASSOCIATED TUBING. THE FSE VERIFIED HARDWARE BY PERFORMING QC, WHICH WAS WITHIN THE ESTABLISHED RANGES. THE FSE ALSO PERFORMED HSFOAM/NOFOAM TEST, AND THE RESULTS WERE WITHIN INSTRUMENT SPECIFICATIONS. HARDWARE WAS THE ROOT CAUSE OF THIS EVENT.
A CUSTOMER CONTACTED BECKMAN COULTER, INC. (BCI) TO REPORT A LEAK COMING FROM UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM. THERE WAS NO EXPOSURE TO THE LEAK AND NO REPORT OF INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXI 800 ACCESS® IMMUNOASSAY ANALYZER | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | JJE | BECKMAN COULTER, INC. | DXI 800 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |