CAPSURE EPI
MDR report key: 2063531
·
Received April 19, 2011
Report
- Report Number
- 2182208-2011-00641
- Event Type
- Injury
- Date Received
- April 19, 2011
- Manufacturer
- MEDTRONIC, INC.
- Product Code
- DTB
- PMA / PMN Number
- P950024/S2
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. IT WAS NOTED THAT ALL CONDUCTORS DISTORTED, THE DISTAL CONDUCTOR WAS STRETCHED, THERE WAS BLOOD/BODY FLUID ON THE DISTAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THE TINES WERE CUT AND THERE WAS APPARENT EXPLANT DAMAGE.THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE LEAD HAD HIGH CAPTURE THRESHOLDS. THE LEAD WAS REMOVED AND SUCCESSFULLY REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CAPSURE EPI | IMPLANTABLE PACING LEAD | DTB | MEDTRONIC, INC. | 4968 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 12 YR | Hospitalization| R | 4968 IMPLANTABLE PACING LEAD| P1501DR IMPLANTABLE PULSE GENERATOR |