2 LEAD SOLUTION ADMINISTRATIONSET
Report
- Report Number
- 6000001-2011-02980
- Event Type
- Malfunction
- Date Received
- April 19, 2011
- Date of Event
- April 13, 2011
- Report Date
- April 13, 2011
- Manufacturer
- BAXTER HEALTHCARE - MALTA
- Product Code
- FPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DA
- Reporter Occupation
- NURSE
Narratives
(B)(4). THE SAMPLE WAS RECEIVED FOR EVALUATION; HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF EVALUATION OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT HAVE A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.
(B)(4). AN ACTUAL SAMPLE WAS RECEIVED FOR EVALUATION. THE SAMPLE WAS SUBMITTED FOR FUNCTIONAL TESTING USING A METHYLENE BLUE SOLUTION TO DETECT ANY LEAKAGE CAUSED BY A FAULTY CLAMP. THIS ANALYSIS CONFIRMED THAT THE CLAMP IS DEFECTIVE. THE REPORTED CONDITION WAS CONFIRMED. ALTHOUGH THE REPORTED CONDITION WAS CONFIRMED, THE ROOT CAUSE WAS NOT IDENTIFIED. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.
THE CUSTOMER REPORTED TO BAXTER (B)(4) A 2-LEAD SOLUTION ADMINISTRATION SET IN WHICH THE CLAMP DID NOT WORK - IT WAS IMPOSSIBLE TO CLOSE. THE CONDITION OCCURRED WHILE ADMINISTERING AN UNKNOWN ANTIBIOTIC ON A PATIENT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 2 LEAD SOLUTION ADMINISTRATIONSET | SET, ADMINISTRATION, INTRAVASCULAR | FPA | BAXTER HEALTHCARE - MALTA | 10J05V067M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN ANTIBIOTIC |