FDA Adverse Event Malfunction Summary report: N

2 LEAD SOLUTION ADMINISTRATIONSET

MDR report key: 2063515 · Received April 19, 2011

Report

Report Number
6000001-2011-02980
Event Type
Malfunction
Date Received
April 19, 2011
Date of Event
April 13, 2011
Report Date
April 13, 2011
Manufacturer
BAXTER HEALTHCARE - MALTA
Product Code
FPA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS RECEIVED FOR EVALUATION; HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF EVALUATION OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT HAVE A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.

Additional Manufacturer Narrative · 1

(B)(4). AN ACTUAL SAMPLE WAS RECEIVED FOR EVALUATION. THE SAMPLE WAS SUBMITTED FOR FUNCTIONAL TESTING USING A METHYLENE BLUE SOLUTION TO DETECT ANY LEAKAGE CAUSED BY A FAULTY CLAMP. THIS ANALYSIS CONFIRMED THAT THE CLAMP IS DEFECTIVE. THE REPORTED CONDITION WAS CONFIRMED. ALTHOUGH THE REPORTED CONDITION WAS CONFIRMED, THE ROOT CAUSE WAS NOT IDENTIFIED. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO BAXTER (B)(4) A 2-LEAD SOLUTION ADMINISTRATION SET IN WHICH THE CLAMP DID NOT WORK - IT WAS IMPOSSIBLE TO CLOSE. THE CONDITION OCCURRED WHILE ADMINISTERING AN UNKNOWN ANTIBIOTIC ON A PATIENT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 2 LEAD SOLUTION ADMINISTRATIONSET SET, ADMINISTRATION, INTRAVASCULAR FPA BAXTER HEALTHCARE - MALTA 10J05V067M

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN ANTIBIOTIC