FDA Adverse Event Malfunction Summary report: N

IMPLANTABLE DRUG INFUSION PUMP

MDR report key: 20634435 · Received November 7, 2024

Report

Report Number
2182207-2024-04503
Event Type
Malfunction
Date Received
November 7, 2024
Date of Event
October 10, 2024
Report Date
April 4, 2025
Manufacturer
MEDTRONIC NEUROMODULATION
Product Code
LKK
PMA / PMN Number
P860004
Removal / Correction Number
Z-2352-2024
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

G4 CORRECTION/UPDATE: THE INITIAL REPORT DID NOT INCLUDE PMA# P860004 IN ERROR. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: PRODUCT ID 8781, LOT# N772820001, IMPLANTED: (B)(6) 2018, PARTIALLY EXPLANTED: (B)(6) 2024, PRODUCT TYPE: CATHETER. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: PRODUCT ID 8781, LOT# N772820001, CATHETER. SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 8781, SERIAL/LOT #: (B)(6), UBD: 07-DEC-2019, UDI#: (B)(4), MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

ADDITIONAL INFORMATION WAS RECEIVED FROM A DISTRIBUTOR VIA A HEALTHCARE PROVIDER (FOREIGN). AN ASCENDA CATHETER WAS OPENED FOR PARTIAL REPLACEMENT.

Description of Event or Problem · 0

ADDITIONAL INFORMATION WAS RECEIVED FROM A FOREIGN HEALTHCARE PROVIDER VIA A DISTRIBUTOR. THE MODEL NUMBER, SERIAL NUMBER, AND IMPLANT DATE OF THE PUMP WAS UNKNOWN. REGARDING THE INITIAL REASON FOR THE PROCEDURE IN WHICH THE ISSUE WITH THE CONNECTOR WAS IDENTIFIED, TISSUE INSERTION WAS INDICATED. ONLY REPLACEMENT (PARTIAL REPLACEMENT) OF THE CATHETER OCCURRED. THE EXPLANTED CATHETER COMPONENT WAS DISCARDED BY THE HEALTHCARE PROVIDER.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM A DISTRIBUTOR VIA A HEALTHCARE PROVIDER (FOREIGN) REGARDING A PATIENT RECEIVING AN UNKNOWN DRUG OF AN UNKNOWN CONCENTRATION AT AN UNKNOWN DOSE RATE. IT WAS REPORTED THAT A CATHETER CONNECTOR DEFECT OCCURRED. WHEN THE CATHETER SUTURELESS PUMP CONNECTOR WAS REMOVED FROM THE SYNCHROMED II PUMP CATHETER PORT, TISSUE WAS FOUND TO HAVE INFILTRATED INTO THE CONNECTOR PART. THE TISSUE FRAGMENTS WERE REMOVED BY VISUAL INSPECTION, BUT THE OUTER CYLINDER OF AN 18G SURFLOW NEEDLE FOR REMOVING THE DRUG SOLUTION COULD NOT BE INSERTED. NO PATIENT SYMPTOM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1903722 IMPLANTABLE DRUG INFUSION PUMP PUMP, INFUSION, IMPLANTED, PROGRAMMABLE LKK MEDTRONIC NEUROMODULATION NEU_UNKNOWN_PUMP

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown SEE H11....