FDA Adverse Event Malfunction Summary report: N

DIMENSION® CLINICAL CHEMISTRY SYSTEM

MDR report key: 2063367 · Received April 19, 2011

Report

Report Number
1226181-2011-00036
Event Type
Malfunction
Date Received
April 19, 2011
Date of Event
April 5, 2011
Report Date
April 5, 2011
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC - BROOKFIELD
Product Code
JJE
PMA / PMN Number
K944093
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF THE INSTRUMENT AND INSTRUMENT DATA INDICATE THAT THE CAUSE FOR THE FALSELY ELEVATED TROPONIN I RESULT WAS HM CHROME CONTAMINATION. A SIEMENS HEALTHCARE DIAGNOSTICS INC FIELD SERVICE REPRESENTATIVE (FSE) WAS DISPATCHED TO THE ACCOUNT TO PERFORM MAINTENANCE PROCEDURES. THE INSTRUMENT IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

A FALSELY ELEVATED TROPONIN I RESULT WAS OBTAINED ON A PATIENT SAMPLE. THE RESULT WAS NOT REPORTED TO THE PHYSICIAN. THE SAMPLE WAS REPEATED AND A NEGATIVE RESULT WAS OBTAINED. PATIENT TREATMENT WAS NOT ALTERED OR PRESCRIBED. THERE WAS NO REPORT OF ADVERSE HEALTH CONSEQUENCES AS A RESULT OF THE FALSELY ELEVATED TROPONIN I RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DIMENSION® CLINICAL CHEMISTRY SYSTEM CLINICAL CHEMISTRY SYSTEM JJE SIEMENS HEALTHCARE DIAGNOSTICS INC - BROOKFIELD RXL MAX HM

Patients

Seq Age Sex Outcome Treatment
1