STERLING ES PTA BALLOON DILATATION CATHETER

MDR report key: 2063310 · Received April 19, 2011

Report

Report Number
2134265-2011-01156
Event Type
Malfunction
Date Received
April 19, 2011
Date of Event
March 24, 2011
Report Date
March 24, 2011
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
LIT
PMA / PMN Number
K080982
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE COMPLAINT DEVICE WAS NOT RECEIVED FOR ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW FOR THE REPORTED BATCH CONFIRMED THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CORONARY ARTERY TREATMENT PROCEDURE, A BALLOON RUPTURE OCCURRED. THE 100% STENOSED, SEVERELY TORTUOUS, MODERATELY CALCIFIED LESION WAS LOCATED IN POSTERIOR TIBIAL ARTERY. UPON 1ST INFLATION, THE STERLING 1.5MM X 20MM X 143CM BALLOON RUPTURED AT 6ATM. THE DEVICE WAS REMOVED, AND THE PROCEDURE COMPLETED WITH A DIFFERENT DEVICE. THERE WERE NO REPORTED COMPLICATIONS AND THE PATIENT CONDITION IS GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STERLING ES PTA BALLOON DILATATION CATHETER CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL LIT BOSTON SCIENTIFIC - MAPLE GROVE H74939135152010 14047236

Patients

Seq Age Sex Outcome Treatment
1 CRUISE SJM GUIDEWIRE| 4.5F PARENT SHEATH MEDIKIT INTRODUCER SHEATH