RADIFOCUS GLIDEWIRE ADVANTAGE
Report
- Report Number
- 9681834-2011-00015
- Event Type
- Malfunction
- Date Received
- April 13, 2011
- Date of Event
- March 15, 2011
- Report Date
- March 15, 2011
- Manufacturer
- TERUMO CORPORATION, ASHITAKA
- Product Code
- DQX
- PMA / PMN Number
- K063372
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
RESULTS: IS BASED UPON EVAL OF USER FACILITY INFO AND RETURNED SAMPLE; IS BASED UPON EVAL OF UNDAMAGED SECTIONS OF THE RETURNED SAMPLE. CONCLUSIONS: IS BASED UPON EVAL OF USER FACILITY INFO AND RETURNED SAMPLE; IS BASED UPON EVAL OF UNDAMAGED SECTIONS OF THE RETURNED SAMPLE. THE INVOLVED DEVICE WAS RETURNED AND EVALUATED BY QA AT THE MFG FACILITY. EXAMINATION CONFIRMED: THE URETHANE COATING HAD BEEN PEELED AWAY FROM THE CORE WIRE STARTING APPROX 250-MM FROM THE DISTAL END; THE COATING WAS THEN ROLLED BACK ON ITSELF IN THE DISTAL DIRECTION EXPOSING APPROX 8-MM OF CORE WIRE AND MEASUREMENT CONFIRMED NONE OF THE URETHANE COATING HAD COMPLETELY DETACHED. INSPECTION AND TESTING OF UNDAMAGED SECTIONS OF THE RETURNED SAMPLE CONFIRMED THAT THERE WERE NO PRE-EXISTING DEFECTS OR ANOMALIES AND PRODUCT PERFORMANCE SPECIFICATIONS WERE MET. A REVIEW OF THE DEVICE HISTORY RECORD INDICATED THAT THERE WERE NO PRODUCTION RELATED PROBLEMS. A REVIEW OF THE COMPLAINT FILES CONFIRMED THAT THIS LOT NUMBER HAS NOT BEEN REPORTED PREVIOUSLY. THE CAUSE FOR THE REPORTED EVENT CANNOT BE DEFINITIVELY DETERMINED BASED ON THE AVAILABLE INFO. THE RESULTS OF THE EVAL OF THE RETURNED SAMPLE ARE MOST CONSISTENT WITH DAMAGE TO THE GUIDEWIRE COATING DUE TO THE DISTAL EDGE OF ANOTHER DEVICE PUSHING BACK THE PROXIMAL END OF THE URETHANE COATING SUCH THAT IT WAS PARTIALLY PEELED OFF THE WIRE AND THEN ROLLED BACK ON ITSELF IN THE DISTAL DIRECTION DURING REMOVAL. THE DEVICE LABELING STATES IN THE WARNINGS/PRECAUTIONS SECTION OF THE INSTRUCTIONS-FOR-USE THAT INAPPROPRIATE MANIPULATION OF THE GLIDEWIRE UNDER CERTAIN CONDITIONS MAY RESULT IN "SHEARING OF THE GLIDEWIRE ADVANTAGE, DESTRUCTION AND/OR SEPARATION OF THE OUTER POLYURETHANE COATING REQUIRING RETRIEVAL." IN ADDITION, THE IFU STATES, "WHEN USING A DRUG OR A DEVICE CONCURRENTLY WITH THE GLIDEWIRE ADVANCE, THE OPERATOR SHOULD HAVE A FULL UNDERSTANDING OF THE PROPERTIES/CHARACTERISTICS OF THE DRUG OR DEVICE SO AS TO AVOID DAMAGE TO THE GLIDEWIRE ADVANTAGE." ALL AVAILABLE INFO HAS BEEN PLACED ON FILE IN QA AT THE MFG FACILITY FOR APPROPRIATE TRACKING, TRENDING AND F/U. (B)(4).
THE USER FACILITY REPORTED THAT A PORTION OF THE GUIDEWIRE COATING HAD BEEN DISPLACED DURING A PROCEDURE EXPOSING, THE CORE WIRE. F/U COMMUNICATION CONFIRMED: RESISTANCE WAS ENCOUNTERED DURING ADVANCING OF THE GUIDEWIRE; THE ENTIRE DEVICE WAS ABLE TO BE REMOVED; THE GLIDEWIRE COATING WAS DESCRIBED AS "PROTRUDING" AT THE DISTAL END OF THE DEVICE AND THERE WAS NO REPORTED PT COMPLICATIONS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RADIFOCUS GLIDEWIRE ADVANTAGE | DQX | TERUMO CORPORATION, ASHITAKA | NA | 101202 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | STENT| 5FR. CORDIS TEMPO CATHETER| 4FR. CORDIS MP CATHETER| 7FR. PTA BALLOON| CORDIS 7FR. POWER FLEX PTA BALLOON |