FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION TREON TREATMENT GUIDANCE SYSTEM

MDR report key: 2063280 · Received March 22, 2011

Report

Report Number
1723170-2011-00699
Event Type
Malfunction
Date Received
March 22, 2011
Date of Event
February 25, 2011
Report Date
February 25, 2011
Manufacturer
MEDTRONIC NAVIGATION, INC.
Product Code
HAW
PMA / PMN Number
K050438
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PT INFO NOT PROVIDED AS NO PT WAS INVOLVED IN THIS CONCERN. PART RETURNED UNDER RMA. THERE IS A BURNT SPOT AT THE +12 VOLTS PIN ON THE CONNECTOR THAT ATTACHES TO THE MONITOR POWER SUPPLY; [SHORT] DIRECTLY CAUSED EVENT.

Description of Event or Problem · 1

A SITE REP REPORTED THE MONITOR DISPLAYED BLACK UPON BOOT-UP AND WAS MAKING A CLICKING SOUND. A MEDTRONIC REP WALKED THE SITE THROUGH TROUBLESHOOTING AND THEY WERE ABLE TO RESTORE THE CONNECTION AFTER RE-SEATING THE CABLE, HOWEVER, THE SAME BEHAVIOR REOCCURRED WITH THE MONITOR GOING BLACK AND CLICKING. THERE WAS NO PT PRESENT.