FDA Adverse Event
Malfunction
Summary report: N
STEALTHSTATION TREON TREATMENT GUIDANCE SYSTEM
MDR report key: 2063280
·
Received March 22, 2011
Report
- Report Number
- 1723170-2011-00699
- Event Type
- Malfunction
- Date Received
- March 22, 2011
- Date of Event
- February 25, 2011
- Report Date
- February 25, 2011
- Manufacturer
- MEDTRONIC NAVIGATION, INC.
- Product Code
- HAW
- PMA / PMN Number
- K050438
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
PT INFO NOT PROVIDED AS NO PT WAS INVOLVED IN THIS CONCERN. PART RETURNED UNDER RMA. THERE IS A BURNT SPOT AT THE +12 VOLTS PIN ON THE CONNECTOR THAT ATTACHES TO THE MONITOR POWER SUPPLY; [SHORT] DIRECTLY CAUSED EVENT.
Description of Event or Problem · 1
A SITE REP REPORTED THE MONITOR DISPLAYED BLACK UPON BOOT-UP AND WAS MAKING A CLICKING SOUND. A MEDTRONIC REP WALKED THE SITE THROUGH TROUBLESHOOTING AND THEY WERE ABLE TO RESTORE THE CONNECTION AFTER RE-SEATING THE CABLE, HOWEVER, THE SAME BEHAVIOR REOCCURRED WITH THE MONITOR GOING BLACK AND CLICKING. THERE WAS NO PT PRESENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STEALTHSTATION TREON TREATMENT GUIDANCE SYSTEM | STEREOTAXIC INSTRUMENT (HAW) | HAW | MEDTRONIC NAVIGATION, INC. | TREON | NA |