EQUINOXE REVERSE SHOULDER COMPONENTS
Report
- Report Number
- 1038671-2024-04304
- Event Type
- Injury
- Date Received
- November 7, 2024
- Date of Event
- October 18, 2024
- Report Date
- October 28, 2025
- Manufacturer
- EXACTECH, INC.
- Product Code
- KWT
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
PENDING INVESTIGATION. D10: 320-10-05 EQUINOXE REVERSE TRAY ADAPTER PLATE +5MM 7203013; 320-38-03 145 DEGREE 38MM HUMERAL LINER +2.5 S395220; 300-01-10 EQUINOXE HUMERAL STEM 10MM PRIMARY 6007824; 320-15-02 RS GLENOID PLATE SUP. AUG 10 DEG. A188306; 320-06-38 EQUINOXE 38MM GLENOSHERE; 320-20-30 EQ REV COMPRESSION SCREW W/ CAP 4.5 X 30MM A118678; 320-20-30 EQ REV COMPRESSION SCREW W/ CAP 4.5 X 30MM S375299; 320-15-05 EQ REV LOCKING SCREW A218278.
AS REPORTED, APPROXIMATELY 1 MONTH AND 9 DAYS POST THE FIRST REVISION SURGERY, THE PATIENT PRESENTED BACK WITH CONTINUED ISSUES WITH INSTABILITY AND PAIN. THE PATIENT UNDERWENT A SECOND REVISION. THE SURGEON HAD ISSUES WITH BEING ABLE TO GET THE TORQUE SCREW OUT OF THE HUMERAL STEM AND TRAY. IT WAS NOT FULLY RELEASED FROM THE STEM, CAUSING THE HUMERAL TRAY TO SPIN CIRCUMFERENTIALLY AROUND THE STEM. THE SURGEON REPLACED THE GLENOSPHERE WITH A LATERALIZED ONE, KEPT THE TRAY AS IT WOULD NOT RELEASE FROM THE STEM, AND IMPLANTED A CONSTRAINED LINER. THERE WAS A 5-15 MINUTE DELAY. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EVENTS AS A RESULT OF THE DELAY. THE SURGEON FELT HE GOT A STABLE CONSTRUCT WITH THE IMPLANTS THAT HE WAS ABLE TO EXCHANGE. HE WOULD HAVE LIKED TO BE ABLE TO RELEASE THE HUMERAL TRAY FROM THE HUMERAL STEM AND SPENT SOME TIME TRYING TO DISENGAGE THE IMPLANTS HOWEVER WAS NOT ABLE TO GET THEM TO DISENGAGE. THE PATIENT WAS LAST KNOWN TO BE IN STABLE CONDITION FOLLOWING THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1796820 | EQUINOXE REVERSE SHOULDER COMPONENTS | PROSTHESIS, SHOULDER, NON-CONSTRAINED, METAL/POLYMER CEMENTED | KWT | EXACTECH, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Female | Required Intervention |