FDA Adverse Event Injury Summary report: N

STARCLOSE SE VASCULAR CLOSURE SYSTEM

MDR report key: 2063236 · Received April 19, 2011

Report

Report Number
2024168-2011-02738
Event Type
Injury
Date Received
April 19, 2011
Date of Event
March 17, 2011
Report Date
March 25, 2011
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P050007
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. A DEVICE CAN BE DIFFICULT TO REMOVE DUE TO A NUMBER OF FACTORS INCLUDING, BUT NOT ARE LIMITED TO, TISSUE COMPACTION THAT RESULTS IN A DISTAL FORCE BEING APPLIED TO THE LOCATOR WINGS, BENDING THEM DISTALLY, AND RESTRICTING THEIR PROPER RETRACTION INTO THE DELIVERY TUBESET. ULTIMATELY, THE RETURN OF THE DEVICE MAY HAVE AIDED THE INVESTIGATION IN DETERMINING A CAUSE FOR THE EXPERIENCED EVENT. TO ENSURE THIS TYPE OF EXPERIENCE IS NOT A RESULT OF A POTENTIAL PRODUCT RELATED DEFICIENCY, A SAMPLING OF FINISHED DEVICES ARE TESTED TO VERIFY THE FUNCTIONALITY OF THE DEVICE. A REVIEW OF THE FINISHED PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY MANUFACTURING RELATED NON-CONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT. IN ADDITION, A QUERY OF THE COMPLAINT-HANDLING DATABASE WAS PERFORMED AND THERE HAVE BEEN NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE INFORMATION RECEIVED WITH THIS INCIDENT AND WITHOUT THE PRODUCT TO EXAMINE, A DEFINITIVE CAUSE FOR THE REPORTED EXPERIENCE CANNOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED THAT PHYSICIANS TRAINED IN THE USE OF THE STARCLOSE SE DEVICE ACHIEVED HEMOSTASIS OF THE COMMON FEMORAL ARTERY AFTER INTERVENTIONAL PROCEDURES. REPORTEDLY, AFTER CLIP DEPLOYMENT, THE DEVICE WAS SLIGHTLY DIFFICULT TO REMOVE FROM THE PATIENT ANATOMY. IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE, THE ACCESS PORTS WERE USED TO SUCCESSFULLY FACILITATE DEVICE REMOVAL. THE CLIP OF THE STARCLOSE SE DEVICE SUCCESSFULLY ACHIEVED HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STARCLOSE SE VASCULAR CLOSURE SYSTEM IMPLANTABLE CLIP MGB AV-TEMECULA-CT 010276H

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention