STARCLOSE SE VASCULAR CLOSURE SYSTEM
Report
- Report Number
- 2024168-2011-02738
- Event Type
- Injury
- Date Received
- April 19, 2011
- Date of Event
- March 17, 2011
- Report Date
- March 25, 2011
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- MGB
- PMA / PMN Number
- P050007
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. A DEVICE CAN BE DIFFICULT TO REMOVE DUE TO A NUMBER OF FACTORS INCLUDING, BUT NOT ARE LIMITED TO, TISSUE COMPACTION THAT RESULTS IN A DISTAL FORCE BEING APPLIED TO THE LOCATOR WINGS, BENDING THEM DISTALLY, AND RESTRICTING THEIR PROPER RETRACTION INTO THE DELIVERY TUBESET. ULTIMATELY, THE RETURN OF THE DEVICE MAY HAVE AIDED THE INVESTIGATION IN DETERMINING A CAUSE FOR THE EXPERIENCED EVENT. TO ENSURE THIS TYPE OF EXPERIENCE IS NOT A RESULT OF A POTENTIAL PRODUCT RELATED DEFICIENCY, A SAMPLING OF FINISHED DEVICES ARE TESTED TO VERIFY THE FUNCTIONALITY OF THE DEVICE. A REVIEW OF THE FINISHED PRODUCT LOT HISTORY RECORD DID NOT REVEAL ANY MANUFACTURING RELATED NON-CONFORMING MATERIAL RECORDS ASSOCIATED WITH THIS LOT. IN ADDITION, A QUERY OF THE COMPLAINT-HANDLING DATABASE WAS PERFORMED AND THERE HAVE BEEN NO OTHER SIMILAR INCIDENTS REPORTED FROM THIS LOT. BASED ON THE INFORMATION RECEIVED WITH THIS INCIDENT AND WITHOUT THE PRODUCT TO EXAMINE, A DEFINITIVE CAUSE FOR THE REPORTED EXPERIENCE CANNOT BE DETERMINED.
IT WAS REPORTED THAT PHYSICIANS TRAINED IN THE USE OF THE STARCLOSE SE DEVICE ACHIEVED HEMOSTASIS OF THE COMMON FEMORAL ARTERY AFTER INTERVENTIONAL PROCEDURES. REPORTEDLY, AFTER CLIP DEPLOYMENT, THE DEVICE WAS SLIGHTLY DIFFICULT TO REMOVE FROM THE PATIENT ANATOMY. IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE, THE ACCESS PORTS WERE USED TO SUCCESSFULLY FACILITATE DEVICE REMOVAL. THE CLIP OF THE STARCLOSE SE DEVICE SUCCESSFULLY ACHIEVED HEMOSTASIS. THERE WERE NO REPORTED ADVERSE PATIENT SEQUELAE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STARCLOSE SE VASCULAR CLOSURE SYSTEM | IMPLANTABLE CLIP | MGB | AV-TEMECULA-CT | 010276H |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |