PCA PLS II
Report
- Report Number
- 9615050-2011-00179
- Event Type
- Malfunction
- Date Received
- March 22, 2011
- Date of Event
- January 1, 2011
- Report Date
- February 24, 2011
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K912928
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
INITIAL TESTING OF THE DEVICE WAS CONDUCTED AT THE USER FACILITY BY THE FIELD SERVICE ENGINEER ON (B)(4) 2011. THE PT PENDANT IS EXPECTED TO BE RETURNED FOR FURTHER TESTING. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL. (B)(4).
THE CUSTOMER CONTACT REPORTED THE DEVICE DID NOT DELIVER A DOSE WHEN THE PT PENDANT WAS PRESSED. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED, "BROKEN". NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, DEVICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT DELIVER A DOSE WHEN THE PT PENDANT WAS PRESSED. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PCA PLS II | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |