FDA Adverse Event Malfunction Summary report: N

PCA PLS II

MDR report key: 2063210 · Received March 22, 2011

Report

Report Number
9615050-2011-00179
Event Type
Malfunction
Date Received
March 22, 2011
Date of Event
January 1, 2011
Report Date
February 24, 2011
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K912928
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

INITIAL TESTING OF THE DEVICE WAS CONDUCTED AT THE USER FACILITY BY THE FIELD SERVICE ENGINEER ON (B)(4) 2011. THE PT PENDANT IS EXPECTED TO BE RETURNED FOR FURTHER TESTING. INVESTIGATION IS NOT COMPLETE. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THE DEVICE DID NOT DELIVER A DOSE WHEN THE PT PENDANT WAS PRESSED. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED, "BROKEN". NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, DEVICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. DURING TESTING AT THE USER FACILITY, THE DEVICE DID NOT DELIVER A DOSE WHEN THE PT PENDANT WAS PRESSED. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PCA PLS II 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 UNK