FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2063207 · Received April 12, 2011

Report

Report Number
2531779-2011-02490
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
March 15, 2011
Report Date
March 15, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION CONFIRMED THAT ALL BUTTON SYMBOLS ARE WORN, AND THAT ALL KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE TO PRESSES. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL BUTTON KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PATIENT REPORTED THAT THE UP ARROW KEYPAD BUTTON IS SLOW TO RESPOND TO PRESSES AT TIMES. SHE NOTED THAT THE BUTTON DOES NOT CLICK WHEN PRESSED. THE PATIENT CONFIRMED THAT THE RUBBER KEYPAD IS INTACT; STATED THAT THE BUTTON SYMBOLS ARE WORN OFF; DENIED EXPOSING THE PUMP TO MOISTURE; AND STATED THAT SHE WEARS THE PUMP CLIPPED TO HER WAIST.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 40 YR