FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2063204 · Received April 12, 2011

Report

Report Number
2531779-2011-02492
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
March 15, 2011
Report Date
March 15, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION CONFIRMED THAT THE UP ARROW KEYPAD BUTTON REQUIRES MULTIPLE PRESSES TO OBTAIN A RESPONSE. IT WAS OBSERVED THAT ALL BUTTON KEY CONTACTS SPRING BACK WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL BUTTON KEY CONTACTS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PATIENT REPORTED THAT THE UP ARROW KEYPAD BUTTON HAS BEEN INTERMITTENTLY UNRESPONSIVE FOR A FEW WEEKS. SHE CONFIRMED THAT THE RUBBER KEYPAD IS INTACT; DENIED EXPOSING THE PUMP TO WATER AND CLEANING PRODUCTS; AND STATED THAT SHE WEARS THE PUMP IN HER POCKET, IN HER BRA, OR ON HER WAISTBAND.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 71 YR