FDA Adverse Event
Malfunction
Summary report: N
ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM
MDR report key: 2063204
·
Received April 12, 2011
Report
- Report Number
- 2531779-2011-02492
- Event Type
- Malfunction
- Date Received
- April 12, 2011
- Date of Event
- March 15, 2011
- Report Date
- March 15, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION CONFIRMED THAT THE UP ARROW KEYPAD BUTTON REQUIRES MULTIPLE PRESSES TO OBTAIN A RESPONSE. IT WAS OBSERVED THAT ALL BUTTON KEY CONTACTS SPRING BACK WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER ALL BUTTON KEY CONTACTS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PATIENT REPORTED THAT THE UP ARROW KEYPAD BUTTON HAS BEEN INTERMITTENTLY UNRESPONSIVE FOR A FEW WEEKS. SHE CONFIRMED THAT THE RUBBER KEYPAD IS INTACT; DENIED EXPOSING THE PUMP TO WATER AND CLEANING PRODUCTS; AND STATED THAT SHE WEARS THE PUMP IN HER POCKET, IN HER BRA, OR ON HER WAISTBAND.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ONE TOUCH PING |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 71 YR |