FDA Adverse Event Malfunction Summary report: N

ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM

MDR report key: 2063203 · Received April 12, 2011

Report

Report Number
2531779-2011-02495
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
March 16, 2011
Report Date
March 16, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. THE KEYPAD WAS FOUND TO BE INTACT; NO PEELING OR LIFTING WAS OBSERVED. IT WAS CONFIRMED THAT THE BUTTON SYMBOLS ARE WORN. THE UP ARROW KEYPAD BUTTON WAS INTERMITTENTLY RESPONSIVE DURING TESTING. DURING INVESTIGATION, THE UP ARROW KEY CONTACT WAS OBSERVED TO BE MISALIGNED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE KEYPAD BUTTONS WERE INTERMITTENTLY UNRESPONSIVE. THE PATIENT REPORTED THAT THE UP ARROW KEYPAD BUTTON HAS BEEN INTERMITTENTLY UNRESPONSIVE. SHE NOTED THAT THE UP ARROW BUTTON DOES NOT CLICK WHEN PRESSED; THE UP ARROW AND OK KEYPAD BUTTON SYMBOLS ARE WORN; AND THE RUBBER KEYPAD IS WORN THIN. THE PATIENT DENIED EXPOSING THE PUMP TO WATER; AND STATED THAT SHE CARRIES THE PUMP ON A CLIP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONE TOUCH PING GLUCOSE MANAGEMENT SYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORP. ONE TOUCH PING

Patients

Seq Age Sex Outcome Treatment
1 46 YR