FDA Adverse Event Malfunction Summary report: N

M SERIES

MDR report key: 2063155 · Received April 12, 2011

Report

Report Number
1220908-2011-00902
Event Type
Malfunction
Date Received
April 12, 2011
Report Date
March 21, 2011
Manufacturer
ZOLL MEDICAL CORPORATION
Product Code
MKJ
PMA / PMN Number
K990762
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

ZOLL MEDICAL CORP HAS NOT RECEIVED THE DEVICE FOR EVALUATION AND THIS COMPLAINT IS STILL UNDER INVESTIGATION.

Description of Event or Problem · 1

COMPLAINANT ALLEGED THAT DURING TRAINING BY FACILITY STAFF, THE DEVICE'S PACER OUTPUT WAS NOT ADJUSTABLE. COMPLAINANT INDICATED THAT THERE WAS NO PT INVOLVEMENT IN THE REPORTED MALFUNCTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 M SERIES DEFIBRILLATOR/PACEMAKER MKJ ZOLL MEDICAL CORPORATION MSERIES BI-PHASIC NA

Patients

Seq Age Sex Outcome Treatment
1 NA