FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2063146 · Received April 12, 2011

Report

Report Number
2531779-2011-02493
Event Type
Malfunction
Date Received
April 12, 2011
Report Date
March 19, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Removal / Correction Number
2531779-03/24/2010-003-R
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVALUATION. EVALUATION REVEALED A DISLODGED DISPLAY SCREEN AND PARTIALLY DISLODGED FORCE SENSOR PINS. A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH WAS NOT DUPLICATED DURING TESTING.

Description of Event or Problem · 1

PT REPORTS PUMP DISPENSED INSULIN DURING LOAD CARTRIDGE PHASE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1