FDA Adverse Event Malfunction Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 2063117 · Received April 12, 2011

Report

Report Number
2032227-2011-00915
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
April 2, 2011
Report Date
April 2, 2011
Manufacturer
MEDTRONIC MINIMED
Product Code
MDS
PMA / PMN Number
K040676
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER CALLED TO REPORT A MOTOR ERROR ALARM DURING THE REWIND. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE DISPLACEMENT TEST. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC MINIMED MMT-515LNAS

Patients

Seq Age Sex Outcome Treatment
1 56 YR