FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2063101
·
Received April 12, 2011
Report
- Report Number
- 2531779-2011-02466
- Event Type
- Malfunction
- Date Received
- April 12, 2011
- Report Date
- March 15, 2011
- Manufacturer
- ANIMAS CORP.
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. IT WAS ALSO FOUND THAT MULTIPLE PRESSES ARE REQUIRED TO ACTIVATE A RESPONSE. DURING INVESTIGATION, THE UP ARROW KEY CONTACT WAS OBSERVED TO BE MISALIGNED. ALL KEYPAD BUTTONS DID NOT HAVE NORMAL SPRING BACK. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT DURING EVAL, THE DISPLAY WAS FADED UPON START UP, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE OK BUTTON REQUIRES SEVERAL PRESSES FOR A RESPONSE. IT WAS REPORTED THAT THE KEYPAD IS INTACT AND THE PUMP IS NOT EXPOSED TO WATER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORP. | ANIMAS 2020 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |