FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2063101 · Received April 12, 2011

Report

Report Number
2531779-2011-02466
Event Type
Malfunction
Date Received
April 12, 2011
Report Date
March 15, 2011
Manufacturer
ANIMAS CORP.
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS FOR EVAL. ALL KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS DURING TESTING. IT WAS ALSO FOUND THAT MULTIPLE PRESSES ARE REQUIRED TO ACTIVATE A RESPONSE. DURING INVESTIGATION, THE UP ARROW KEY CONTACT WAS OBSERVED TO BE MISALIGNED. ALL KEYPAD BUTTONS DID NOT HAVE NORMAL SPRING BACK. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS. UNRELATED TO THE COMPLAINT DURING EVAL, THE DISPLAY WAS FADED UPON START UP, WHICH HAS NO EFFECT ON INSULIN DELIVERY FUNCTION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE OK BUTTON REQUIRES SEVERAL PRESSES FOR A RESPONSE. IT WAS REPORTED THAT THE KEYPAD IS INTACT AND THE PUMP IS NOT EXPOSED TO WATER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORP. ANIMAS 2020 NA

Patients

Seq Age Sex Outcome Treatment
1