FDA Adverse Event Malfunction Summary report: N

AFFINITY BED

MDR report key: 2063100 · Received March 16, 2011

Report

Report Number
1824206-2011-01624
Event Type
Malfunction
Date Received
March 16, 2011
Date of Event
May 9, 2008
Report Date
May 9, 2008
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THAT THE LEFT CAM PIVOT WAS CRANKED CAUSING THE LEFT FOOT BRAKE CASTER NOT TO ENGAGE IN THE BRAKE MODE. THE TECHNICIAN REPLACED THE CAM PIVOT AND ADJUSTED THE BRAKE CASTER TO RESOLVE THE ISSUE. THIS MDR IS A RESULT OF CAPA ACTIVITY FOR THE RETROSPECTIVE REVIEW OF COMPLAINTS.

Description of Event or Problem · 1

THE TECHNICIAN ALLEGED THAT THE LEFT FOOT BRAKE CASTER DID NOT HOLD WHEN THE BRAKES WERE ENGAGED. NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AFFINITY BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 3700

Patients

Seq Age Sex Outcome Treatment
1