FDA Adverse Event
Malfunction
Summary report: N
AFFINITY BED
MDR report key: 2063100
·
Received March 16, 2011
Report
- Report Number
- 1824206-2011-01624
- Event Type
- Malfunction
- Date Received
- March 16, 2011
- Date of Event
- May 9, 2008
- Report Date
- May 9, 2008
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THAT THE LEFT CAM PIVOT WAS CRANKED CAUSING THE LEFT FOOT BRAKE CASTER NOT TO ENGAGE IN THE BRAKE MODE. THE TECHNICIAN REPLACED THE CAM PIVOT AND ADJUSTED THE BRAKE CASTER TO RESOLVE THE ISSUE. THIS MDR IS A RESULT OF CAPA ACTIVITY FOR THE RETROSPECTIVE REVIEW OF COMPLAINTS.
Description of Event or Problem · 1
THE TECHNICIAN ALLEGED THAT THE LEFT FOOT BRAKE CASTER DID NOT HOLD WHEN THE BRAKES WERE ENGAGED. NO INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AFFINITY BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM, INC. | 3700 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |