FDA Adverse Event Malfunction Summary report: N

ADVANCE BED

MDR report key: 2063095 · Received March 16, 2011

Report

Report Number
1824206-2011-01603
Event Type
Malfunction
Date Received
March 16, 2011
Date of Event
July 29, 2008
Report Date
July 29, 2008
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN REPLACED THE BED EXIT TAPE SWITCHES TO RESOLVE THE ISSUE. THIS MDR IS A RESULT OF CAPA ACTIVITY FOR THE RETROSPECTIVE REVIEW OF COMPLAINTS.

Description of Event or Problem · 1

TECHNICIAN ALLEGED THAT THE BED EXIT IS NOT ALARMING WITH WEIGHT IN THE BED. NO INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANCE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1145

Patients

Seq Age Sex Outcome Treatment
1