FDA Adverse Event Malfunction Summary report: N

EVIS EXERA III VIDEO SYSTEM CENTER

MDR report key: 20630823 · Received November 7, 2024

Report

Report Number
3002808148-2024-10864
Event Type
Malfunction
Date Received
November 7, 2024
Date of Event
October 24, 2024
Report Date
November 20, 2024
Manufacturer
SHIRAKAWA OLYMPUS CO., LTD.
Product Code
FET
UDI-DI
04953170298622
PMA / PMN Number
K131780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WY, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S FINAL INVESTIGATION AND DEVICE EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD FOUND NO DEVIATIONS THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS PRESUMED THAT SERIAL DIGITAL INTERFACE (SDI) VIDEO OUTPUT BECOMES INTERMITTENT WAS CAUSED BY THE FAULTY PRINTED CIRCUIT BOARD. HOWEVER, A DEFINITIVE ROOT CAUSE OF THE REPORTED ISSUE COULD NOT BE DETERMINED. OLYMPUS WILL CONTINUE TO MONITOR THE FIELD PERFORMANCE OF THIS DEVICE.

Additional Manufacturer Narrative · 0

THE DEVICE EVALUATION IS ONGOING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

IT WAS OBSERVED DURING THE DEVICE INSPECTION, THAT THE VIDEO SYSTEM CENTER'S SERIAL DIGITAL INTERFACE (SDI) VIDEO OUTPUT WAS INTERMITTENT DUE TO A FAULTY PRINTED CIRCUIT BOARD. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1611557 EVIS EXERA III VIDEO SYSTEM CENTER VIDEO SYSTEM CENTER FET SHIRAKAWA OLYMPUS CO., LTD. CV-190 04953170298622

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown