SURE-VUE URINE HCG
Report
- Report Number
- 2027969-2011-00766
- Event Type
- Malfunction
- Date Received
- April 11, 2011
- Date of Event
- March 19, 2011
- Report Date
- April 11, 2011
- Manufacturer
- ALERE SAN DIEGO, INC.
- Product Code
- JHI
- PMA / PMN Number
- K993317
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
LOT NUMBER(S): HCG0060079. EXPIRATION DATE(S): 04/2012. CONTROL: 25MIU/ML HCG URINE CONTROL LOT: HCG110105-01; 100MIU/ML HCG URINE CONTROL LOT: HCG100513-01; 278.9IU/ML HCG URINE CONTROL LOT: HCG110124-01. THE RETENTION DEVICES MEET QC SPECIFICATIONS. CORRECT POSITIVE RESULTS WERE OBSERVED WHEN TESTED WITH 25MIU/ML HCG URINE CONTROL AT 3 MINUTES READING TIME. (N=5). THE 100MIU/ML HCG URINE CONTROL YIELDED CLEARLY POSITIVE RESULTS AT 3 MINUTES READING TIME. (N=5). THE 278.9IU/ML HCG URINE CONTROL YIELDED CLEARLY POSITIVE RESULTS AT 3 MINUTES READING TIME. (N=4). CONCLUSION: FROM RETAIN TESTING WITH IN-HOUSE CONTROL, THE CLIENT'S RESULT WAS NOT REPLICATED, AND THE PRODUCT PERFORMED AS EXPECTED. THE RETURN DEVICES MEET QC SPECIFICATIONS. CORRECT POSITIVE RESULTS WERE OBSERVED WHEN TESTED WITH 25MIU/ML HCG URINE CONTROL AT 3 MINUTES READING TIME. (N=5). THE 100MIU/ML HCG URINE CONTROL YIELDED CLEARLY POSITIVE RESULTS AT 3 MINUTES READING TIME. (N=5). THE 278.9 IU/ML HCG URINE CONTROL YIELDED CLEARLY POSITIVE RESULTS AT 3 MINUTES READING TIME. (N=5). CONCLUSION: FROM RETURN TESTING WITH IN-HOUSE CONTROL, THE CLIENT'S RESULT WAS NOT REPLICATED, AND THE PRODUCT PERFORMED AS EXPECTED. CORRECTIVE ACTION NOT REQUIRED AT THIS TIME.
CALLER REPORTED POTENTIAL FALSE NEGATIVE HCG RESULTS. RESULTS AS FOLLOWS: PATIENT CAME INTO THE EMERGENCY DEPARTMENT FOR FLANK PAINS. LAST MENSTRUAL PERIOD: (B)(6) 2011. URINE SPECIMEN COLLECTED VIA CLEAN CATCH MIDSTREAM. URINE SPECIMEN NOT AVAILABLE FOR FURTHER TESTING; WAS THROWN AWAY. CT SCAN OF ABDOMEN REVEALED A RESEMBLANCE OF FETUS AND BETA QUANTITATIVE WAS 22,932 ML/IU.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SURE-VUE URINE HCG | HCG PREGNANCY TEST | JHI | ALERE SAN DIEGO, INC. | FHC-102-OBC513 | HCG0060079 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 31 YR |