ANIMAS INSULIN CARTRIDGE
Report
- Report Number
- 2531779-2011-02488
- Event Type
- Malfunction
- Date Received
- April 11, 2011
- Date of Event
- March 13, 2011
- Report Date
- March 13, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K032257
- Removal / Correction Number
- 2531779-02/25/11-001-R
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- OTHER
Narratives
THE CARTRIDGE WAS NOT RETURNED TO ANIMAS. THERE IS NO INVESTIGATION NECESSARY. CARTRIDGES WITH LOT# B201576 WERE CONFIRMED TO BE DEFECTIVE VIA RETAINS. WHEN TESTED, FLUID WAS OBSERVED LEAKING FROM THE PLUNGER END OF THE CARTRIDGE.
THE REPORTER CLAIMED THAT THE PATIENT'S BLOOD GLUCOSE ELEVATED FROM 200-300 MG/DL AND THAT THERE WAS A CARTRIDGE LEAKAGE ISSUE WITH LOT# B201576. THERE WAS NO REPORT OF SYMPTOMS OR MEDICAL INTERVENTION THAT WOULD SUGGEST ACUTE COMPLICATION OF DIABETES AT THE TIME OF CONCERN. IT WAS REPORTED THAT THE PATIENT HAD A COLD. THE PATIENT COULD NOT PROVIDE ANY DETAILS REGARDING DATE OF LEAKAGE OR HOW HIGH BLOOD GLUCOSE CORRELATED TO A CARTRIDGE LEAKAGE AT THE TIME OF CONCERN. DURING TROUBLESHOOTING, THE ANIMAS HEALTHCARE PROVIDER CONCLUDED THAT THE PATIENT'S HIGH BLOOD GLUCOSE MAY HAVE BEEN DUE TO A SITE ISSUE OR NEED FOR CLINICAL INSULIN ADJUSTMENT. THIS COMPLAINT IS BEING REPORTED AS A MALFUNCTION DUE TO THE ALLEGED CARTRIDGE LEAKAGE ISSUE. THE PATIENT'S BLOOD GLUCOSE OF 200-300 MG/DL WITHOUT ANY SYMPTOMS OR MEDICAL INTERVENTION DOES NOT MEET ANIMAS' CRITERIA OF A SERIOUS INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN CARTRIDGE | INSULIN INFUSION PUMP CARTRIDGE | LZG | ANIMAS CORPORATION | IR 1200 / 1250 / 2020/ OTP | B201576 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |