PLUM A+ DRIVER ED 2
Report
- Report Number
- 9615050-2011-00172
- Event Type
- Malfunction
- Date Received
- March 21, 2011
- Date of Event
- February 15, 2011
- Report Date
- February 23, 2011
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K042081
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE IS EXPECTED TO BE RETURNED FOR INVESTIGATION. IT HAS NOT YET BEEN RECEIVED. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL. (B)(4).
THE CUSTOMER CONTACT REPORTED THAT THE PUMP DID NOT SOUND AN AUDIBLE ALARM TONE DURING AN ALARM CONDITION. THE PUMP WAS SENT TO THE BIOMEDICAL ENGINEERING DEPARTMENT WITH AN UNSIGNED NOTE STATING, "ERROR 31." NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. DURING TESTING AT THE USER FACILITY, THE DEVICE DISPLAYED E301 (AUDIO ALARM FAILURE) WITH NO AUDIBLE ALARM TONE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE PUMP WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PLUM A+ DRIVER ED 2 | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | PLUM A+ SOFTWARE MODULE| LIST# 20791, SN (B)(4) |