FDA Adverse Event Malfunction Summary report: N

INTERLINK SYSTEM BASIC SOLNSET INJ SITE

MDR report key: 2062787 · Received April 19, 2011

Report

Report Number
6000001-2011-02966
Event Type
Malfunction
Date Received
April 19, 2011
Date of Event
April 1, 2011
Report Date
April 5, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FPA
PMA / PMN Number
K961225
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS RECEIVED FOR EVALUATION ON 04/28/2011. AN ACTUAL SAMPLE WAS RECEIVED FOR EVALUATION. THE SAMPLE ARRIVED PRIMED AND SPIKED INTO AN EMPTY 2B1322 0.9% SODIUM CHLORIDE SOLUTION BAG. THE SAMPLE WAS REMOVED FROM THE EMPTY BAG AND SPIKED INTO A 1000ML SOLUTION BAG CONTAINING STERILE WATER. THE SAMPLE WAS RE-PRIMED AND NO AIR IN-LINE WAS OBSERVED DURING PRIMING AND FLOW. THE SAMPLE WAS THEN SUBMITTED FOR AN UNDER WATER LEAK TEST AND NO LEAKS WERE OBSERVED. SINCE THE SAMPLE PRIMED AND FLOWED NORMALLY WITH NO LEAKS OR AIR IN-LINE FOUND, THE REPORTED WAS NOT CONFIRMED AND NO ROOT CAUSE WAS IDENTIFIED. A BATCH REVIEW COULD NOT BE COMPLETED AS THE LOT NUMBER WAS NOT PROVIDED BY THE CUSTOMER. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE IS REPORTED TO BE AVAILABLE FOR EVALUATION. IF THE SAMPLE IS RECEIVED OR ADDITIONAL INFORMATION BECOMES AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO BAXTER AN INTERLINK SYSTEM BASIC SOLUTION SET IN WHICH AN AIR IN LINE OCCURRED. ACCORDING TO THE REPORT, THE NURSE WAS ATTEMPTING TO INFUSE VANCOMYCIN ON A PATIENT VIA GRAVITY AND SMALL AIR BUBBLES KEPT APPEARING IN THE TUBING. THE CONDITION OCCURRED DURING INFUSION ON A PATIENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTERLINK SYSTEM BASIC SOLNSET INJ SITE SET, ADMINISTRATION, INTRAVASCULAR FPA BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 VANCOMYCIN