FDA Adverse Event Malfunction Summary report: N

LINE SET FOR AQUARIUS

MDR report key: 2062769 · Received April 19, 2011

Report

Report Number
1423500-2011-04694
Event Type
Malfunction
Date Received
April 19, 2011
Date of Event
April 2, 2011
Report Date
April 3, 2011
Manufacturer
HAEMOTRONICS SPA
Product Code
FJK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS PRODUCT IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT HAVE A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. THE DEVICE HAS BEEN RECEIVED , AND THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION OF THE SAMPLE BY BAXTER AND HAEMOTRONIC CONFIRMED THAT THE PRESSURE DOME WAS DAMAGED AND MISSING 5 TEETH. THE ROOT CAUSE COULD NOT BE DETERMINED DUE TO INSUFFICIENT INFORMATION AVAILABLE. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THE LOT. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED.

Description of Event or Problem · 1

A NURSE REPORTED TO BAXTER (B)(4) THAT THE PRESSURE DOME (RED LINE) WAS DAMAGED WHEN REMOVING IT FROM THE OVERPOUCH. THERE WAS NO PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED ALONG WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LINE SET FOR AQUARIUS SET, TUBING, BLOOD, WITH AND WITHOUT ANTI-REGURGITATION VALVE FJK HAEMOTRONICS SPA B101687

Patients

Seq Age Sex Outcome Treatment
1