INNER SHEATH, FOR 26 FR. OUTER SHEATH
Report
- Report Number
- 9610773-2024-02658
- Event Type
- Malfunction
- Date Received
- November 7, 2024
- Date of Event
- October 17, 2024
- Report Date
- June 11, 2025
- Manufacturer
- OLYMPUS WINTER & IBE GMBH
- Product Code
- HIH
- UDI-DI
- 04042761029339
- PMA / PMN Number
- K931994
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- 003
Narratives
THIS REPORT IS BEING SUPPLEMENTED TO PROVIDE ADDITIONAL INFORMATION BASED ON THE LEGAL MANUFACTURER'S INVESTIGATION. CORRECTED FIELDS: D8 THE CUSTOMER'S ALLEGATION WAS CONFIRMED. THE DEVICE EVALUATION FOUND THE CERAMIC TIP WAS FROM A THIRD PARTY REPAIR. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THAT THE FOLLOWING LED TO THE MALFUNCTION: THIRD PARTY REPAIR SO THAT DEFINED BIOCOMPATIBLE CHARACTERISTICS ARE NO LONGER GUARANTEED (EXAMPLE: IMPROPER ADHESIVES USED; IMPROPER MATERIALS USED) AND DEFINED MECHANICAL CHARACTERISTICS ARE NO LONGER GUARANTEED (MELTING OF IMPROPER INSULATION MATERIAL INSTEAD OF CERAMIC). A DEFINITIVE ROOT CAUSE CANNOT BE IDENTIFIED. A DEVICE HISTORY REVIEW REVEALED NO ISSUES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED ISSUE. OLYMPUS WILL CONTINUE TO MONITOR THE FIELD PERFORMANCE OF THIS DEVICE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
THE DEVICE EVALUATION IS ONGOING. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. G4: PMA/510(K) NUMBER: (K931994/K931995); ADDED HERE DUE TO CHARACTER LIMITATION IN RESPECTIVE FIELD.
IT WAS REPORTED, THE OUTER SHEATH EXHIBITED A FAULTY CERAMIC TIP. THE ISSUE IS UNKNOWN AS TO WHEN IT OCCURRED. THE PATIENT WAS NOT UNDER ANESTHESIA. THERE WERE NO REPORTS OF DELAY, PATIENT HARM, INJURY, OR DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1998850 | INNER SHEATH, FOR 26 FR. OUTER SHEATH | A22040A | HIH | OLYMPUS WINTER & IBE GMBH | A22040A | 131W-0067 | 04042761029339 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |