FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2062332 · Received April 12, 2011

Report

Report Number
1720753-2011-03534
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
March 14, 2011
Report Date
April 12, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE MANUFACTURER'S SERVICE REPRESENTATIVE PERFORMED AN ON-SITE INVESTIGATION. THE GENERATOR WAS CALIBRATED. THE SYSTEM WAS TESTED AND IS OPERATING AS INTENDED. THIS EVENT MAY BE AN ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE KILOVOLTAGE WAS TOO HIGH ON THE 9800 SYSTEM. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1