PROGRAMMING SOFTWARE
Report
- Report Number
- 1644487-2011-00786
- Event Type
- Malfunction
- Date Received
- April 12, 2011
- Date of Event
- August 15, 2005
- Report Date
- March 18, 2011
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
ANALYSIS OF PROGRAMMING HISTORY.
DURING A REVIEW OF A PT'S PROGRAMMING HISTORY AVAILABLE IN THE MANUFACTURER'S PROGRAMMING HISTORY DATABASE, IT WAS FOUND THAT THE PT PRESENTED SETTINGS DIFFERENT THAN WHAT WAS INTENDED DURING IMPLANT ON (B)(6) 2005. THE SETTINGS OCCURRED DUE TO A FAULTED SYSTEMS DIAGNOSTIC TEST ON (B)(6) 2005. AN INTERROGATION OF THE DEVICE WAS PERFORMED, AND THE NORMAL MODE CURRENT SETTING WAS ADJUSTED BACK TO OFF. THE MAGNET MODE CURRENT WAS NOT CORRECTED. AT THE NEXT REPORTED OFFICE VISIT (B)(6) 2006, THE MAGNET CURRENT WAS STILL UNCHANGED. THERE WERE NO REPORTS OF ANY PT ADVERSE EVENTS AS A RESULT. ON (B)(6) 2005 - INTERROGATE - OUTPUT CURRENT: 0 MA, SIGNAL FREQUENCY: 30 HZ, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS, OFF TIME: 180 MINUTES; MAGNET OUTPUT CURRENT: 0 MA, PULSE WIDTH: 500 USEC, ON TIME: 60 SECONDS. ON (B)(6) 2005 - SYSTEMS DIAGNOSTICS - FAULT. ON (B)(6) 2005 - INTERROGATE - OUTPUT CURRENT: 1 MA, SIGNAL FREQUENCY: 20 HZ, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS, OFF TIME: 60 MINUTES, MAGNET OUTPUT CURRENT: 1 MA, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS. ON (B)(6) 2005 - PROGRAM - OUTPUT CURRENT: 0 MA, SIGNAL FREQUENCY: 20 HZ, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS, OFF TIME: 60 MINUTES, MAGNET OUTPUT CURRENT: 1 MA, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS. ON (B)(6) 2005 - INTERROGATE - OUTPUT CURRENT: 0 MA, SIGNAL FREQUENCY: 20 HZ, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS, OFF TIME: 60 MINUTES, MAGNET OUTPUT CURRENT: 1 MA, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS. ON (B)(6) 2006 - INTERROGATE - OUTPUT CURRENT: 1 MA, SIGNAL FREQUENCY: 20 HZ, PULSE WIDTH: 500 USEC, ON TIME: 60 SECONDS, OFF TIME: 60 MINUTES, MAGNET OUTPUT CURRENT: 1 MA, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS. ON (B)(6) 2006 - PROGRAM - OUTPUT CURRENT: 1.25 MA, SIGNAL FREQUENCY: 20 HZ, PULSE WIDTH: 500 USEC, ON TIME: 60 SECONDS, OFF TIME: 60 MINUTES, MAGNET OUTPUT CURRENT: 1 MA, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROGRAMMING SOFTWARE | LYJ | CYBERONICS, INC. | MODEL 250 | 366954 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 28 YR |