FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 2062321 · Received April 12, 2011

Report

Report Number
1644487-2011-00786
Event Type
Malfunction
Date Received
April 12, 2011
Date of Event
August 15, 2005
Report Date
March 18, 2011
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ANALYSIS OF PROGRAMMING HISTORY.

Description of Event or Problem · 1

DURING A REVIEW OF A PT'S PROGRAMMING HISTORY AVAILABLE IN THE MANUFACTURER'S PROGRAMMING HISTORY DATABASE, IT WAS FOUND THAT THE PT PRESENTED SETTINGS DIFFERENT THAN WHAT WAS INTENDED DURING IMPLANT ON (B)(6) 2005. THE SETTINGS OCCURRED DUE TO A FAULTED SYSTEMS DIAGNOSTIC TEST ON (B)(6) 2005. AN INTERROGATION OF THE DEVICE WAS PERFORMED, AND THE NORMAL MODE CURRENT SETTING WAS ADJUSTED BACK TO OFF. THE MAGNET MODE CURRENT WAS NOT CORRECTED. AT THE NEXT REPORTED OFFICE VISIT (B)(6) 2006, THE MAGNET CURRENT WAS STILL UNCHANGED. THERE WERE NO REPORTS OF ANY PT ADVERSE EVENTS AS A RESULT. ON (B)(6) 2005 - INTERROGATE - OUTPUT CURRENT: 0 MA, SIGNAL FREQUENCY: 30 HZ, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS, OFF TIME: 180 MINUTES; MAGNET OUTPUT CURRENT: 0 MA, PULSE WIDTH: 500 USEC, ON TIME: 60 SECONDS. ON (B)(6) 2005 - SYSTEMS DIAGNOSTICS - FAULT. ON (B)(6) 2005 - INTERROGATE - OUTPUT CURRENT: 1 MA, SIGNAL FREQUENCY: 20 HZ, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS, OFF TIME: 60 MINUTES, MAGNET OUTPUT CURRENT: 1 MA, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS. ON (B)(6) 2005 - PROGRAM - OUTPUT CURRENT: 0 MA, SIGNAL FREQUENCY: 20 HZ, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS, OFF TIME: 60 MINUTES, MAGNET OUTPUT CURRENT: 1 MA, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS. ON (B)(6) 2005 - INTERROGATE - OUTPUT CURRENT: 0 MA, SIGNAL FREQUENCY: 20 HZ, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS, OFF TIME: 60 MINUTES, MAGNET OUTPUT CURRENT: 1 MA, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS. ON (B)(6) 2006 - INTERROGATE - OUTPUT CURRENT: 1 MA, SIGNAL FREQUENCY: 20 HZ, PULSE WIDTH: 500 USEC, ON TIME: 60 SECONDS, OFF TIME: 60 MINUTES, MAGNET OUTPUT CURRENT: 1 MA, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS. ON (B)(6) 2006 - PROGRAM - OUTPUT CURRENT: 1.25 MA, SIGNAL FREQUENCY: 20 HZ, PULSE WIDTH: 500 USEC, ON TIME: 60 SECONDS, OFF TIME: 60 MINUTES, MAGNET OUTPUT CURRENT: 1 MA, PULSE WIDTH: 500 USEC, ON TIME: 30 SECONDS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. MODEL 250 366954

Patients

Seq Age Sex Outcome Treatment
1 28 YR