FDA Adverse Event
Malfunction
Summary report: N
VASOVIEW HEMOPRO EVH SYSTEM
MDR report key: 2062309
·
Received March 18, 2011
Report
- Report Number
- 2242352-2011-00160
- Event Type
- Malfunction
- Date Received
- March 18, 2011
- Date of Event
- February 17, 2011
- Report Date
- February 18, 2011
- Manufacturer
- MAQUET CARDIOVASCULAR, LLC
- Product Code
- GEI
- PMA / PMN Number
- K052274
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
INVESTIGATION: A VISUAL INSPECTION REVEALED THAT THE C-RING WAS SPLIT IN HALF. THE OTHER HALF OF THE C-RING AND THE ENTIRE SCOPE WASH TUBING WERE RECEIVED SEPARATE. THERE WAS SOME EVIDENCE OF BLOOD. BASED UPON THE VISUAL OBSERVATIONS, THE REPORTED COMPLAINT FOR "C-RING SPLIT" WAS CONFIRMED. A LOT HISTORY RECORD REVIEW WAS COMPLETED FOR THE PRODUCT. THERE WAS NO NONCONFORMANCE WHICH COULD HAVE CONTRIBUTED TO THIS FAILURE MODE. (B)(4).
Description of Event or Problem · 1
THE HOSP REPORTED THAT DURING AN ENDOSCOPIC VEIN HARVESTING PROCEDURE, THE VASOVIEW HEMOPRO C-RING SPLIT IN HALF DURING THE PROCEDURE. A REPLACEMENT UNIT WAS USED TO COMPLETE THE PROCEDURE. THE HOSP DID NOT REPORT ANY PT EFFECTS. THE PRODUCT WAS RETURNED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VASOVIEW HEMOPRO EVH SYSTEM | ENDOSCOPIC VESSEL HARVESTING SYSTEM | GEI | MAQUET CARDIOVASCULAR, LLC | VH-3000 | 25026693 |