FDA Adverse Event Malfunction Summary report: N

PROGRAMMING SOFTWARE

MDR report key: 2062213 · Received April 11, 2011

Report

Report Number
1644487-2011-00771
Event Type
Malfunction
Date Received
April 11, 2011
Date of Event
March 15, 2011
Report Date
March 15, 2011
Manufacturer
CYBERONICS, INC.
Product Code
LYJ
PMA / PMN Number
P970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE FAILURE OCCURRED, BUT DID NOT CAUSE OR CONTRIBUTE TO A DEATH OR SERIOUS INJURY.

Description of Event or Problem · 1

IT WAS INITIALLY REPORTED THAT A PHYSICIAN'S (B)(4) X50 HANDHELD DEVICE WOULD NOT HOLD A CHARGE. A COMPANY REP VISITED THE SITE TO PERFORM TROUBLESHOOTING AND FOUND THAT THE CONNECTOR TO THE HANDHELD APPEARED TO BE DAMAGED AS THE HANDHELD WOULD ONLY CHARGE WHEN THE CORD WAS HELD A CERTAIN WAY. THERE WAS NO EVIDENCE OF THE BATTERY ON THE HANDHELD WAS SWOLLEN AND THE HANDHELD IS TYPICALLY STORED ON A DESK WHEN NOT IN USE. A REPLACEMENT HANDHELD WAS SENT TO THE PHYSICIAN AND THE DELL X50 HANDHELD AND FLASHCARD WERE RETURNED TO THE MFR FOR ANALYSIS. NO ANOMALIES ASSOCIATED WITH FLASHCARD SOFTWARE OR DATABASES WERE IDENTIFIED DURING THE FLASHCARD ANALYSIS. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO FUNCTIONAL SPECS. DURING ANALYSIS OF THE HANDHELD, IT WAS IDENTIFIED THAT THE SYNC CONNECTOR ON THE BOTTOM OF THE HANDHELD WAS DAMAGED. THE CAUSE FOR THE DAMAGE TO THE CONNECTOR COULD NOT BE DETERMINED BASED ON THE RETURNED HANDHELD. HOWEVER, BASED ON THE DAMAGED CONNECTOR, IT IS MOST LIKELY ASSOCIATED WITH MISHANDLING OF THE DEVICE AS IT APPEARS THE CONNECTOR DETENTS ARE NOT BEING COMPRESSED WHEN REMOVING OR MANIPULATING THE SERIAL DATA CABLE, THUS PLACING AN EXTENSIVE AMOUNT OF FORCE ON THE PCB AND ATTACHED CABLE RECEPTACLE. NO OTHER ANOMALIES WERE IDENTIFIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PROGRAMMING SOFTWARE LYJ CYBERONICS, INC. 250 834042

Patients

Seq Age Sex Outcome Treatment
1